ULTRAFLOW® 54/34
78
Kamstrup A/S · Technical description · 5512-2464_A1_GB_08.2019
9.5.5
Spare parts
Description
Type No.
Battery D-cell
1606-064
Cable retainer (secures the battery)
1650-099
2-pole plug (female)
1643-185
3-pole plug (female)
1643-187
Base PCB (66-R)
5550-517
Table 34. Spare parts for Pulse Tester.
9.5.6
Battery replacement
If Pulse Tester is used continuously, it is recommended to replace the battery once a year.
The battery connector is cut and the cable insulation is removed before the battery is connected to the terminals marked
“Batt”, the red wire to + and the black
wire to -.
Current consumption:
Current consumption with no sensors connected
400
A
Max current consumption with two ULTRAFLOW® connected
1.5 mA
Note
: If the base unit is fitted with battery or externally supplied, the Pulse Tester’s integral supply must be disconnected
(unplugged).
9.6
NOWA
NOWA describes a standardized interface between the heat meter laboratory test equipment (e.g. flow bench) and is
mainly used in laboratories in Germany and Austria. The NOWA test of ULTRAFLOW® 54 (q
p
0.6…100 m³/h)
is supported
in combination with MULTICAL®. For more information about the test of ULTRAFLOW® with NOWA, see Kamstrup
documentation 5585-703 (Hardware) and 5585-706 (Software). NOWA does not support the testing of ULTRAFLOW® 34.
Verification of ULTRAFLOW® 34 is carried out via tracing of pulses.