ULTRAFLOW® 54/34
64
Kamstrup A/S · Technical description · 5512-2464_A1_GB_08.2019
8.8
Interface connector/serial data
ULTRAFLOW® 54 and 34 are fitted with a four-pole connector under the cover. Thus, it is not possible to access the
connector without breaking the seal. The cover is supplied with a factory seal and in connection with verified sensors, it
will be a laboratory seal (legal seal).
The connector is used for:
•
Meter programming, including adjustment of flow charts by means of METERTOOL
•
Bringing the sensor into test mode
•
Reading accumulated water quantity in connection with calibration
•
External control of start/stop in connection with calibration
Because of different physical PCB layouts, the connecting cables of the interfaces in ULTRAFLOW® 54 (type 65-5-XXHX-
XXX) are slightly different from those of other types of ULTRAFLOW® 54 and ULTRAFLOW® 34. The interface connector is
constructed as shown in
. Note that the general functions of the connection are identical, but the location of
the individual functions is different.
(a)
(b)
Figure 54. (a) Interface connector for ULTRAFLOW® 54 (type 65-5-XXHX-XXX) and
(b) ULTRAFLOW® 54 (type 65-5-XXAX-XXX, types 65-5-XXCX-XXX and 65-5-XXJX-XXX) and ULTRAFLOW® 34.
Interface connector (H)
Pin 1 GND
Pin 2 VCC
Pin 3 Pulse out
Pin 4 Verification/Adjustment
Interface connector (A/C/J)
Pin 1 VCC
Pin 2 GND
Pin 3 Pulse out
Pin 4 Verification/Adjustment