Kamoer DIPump550 User Manual Download Page 7

Working mode setting—Flow Mode 

 

 

Fully automatic cycle mode setting 

1. In the flow mode, long press the encoder to switch to the operating flow setting. 

The  nixie  tube  displays  all  numbers  and  flickers.  At  this  time,  it  is  to  set  the 
operating flow of the pump; 

2. The rotary encoder can set the Operating flow, and the setting range is 0 ~ 9999 ml. 
3. Press the encoder (not long press) to switch to the pump stop time setting, and 

the first position of the digital tube will jump   (red in the picture); 

4.  The  rotary  encoder  can  set  the  time,  and  the  settable  time  range  is  0  to  999 

seconds; 

5. Press and hold the encoder to return to the flow rate display interface, and the 

running flow and stop time will be saved automatically; 

6.  Press  the  encoder  (not  long  press),  the  pump  will  run  according  to  the  set 

Operating flow and stop time; 

7. Press the encoder (not long press) and the pump stops cycling. The digital tube 

switches to the display flow rate interface. 

 

Semi-automatic cycle mode setting

 

Simply set the time to 0 in step 4 of the fully automatic cycle mode setting mode. At 
this time, after the pump stops running according to the set Operating flow, it will 
not start automatically; if it starts again, the user needs to press the encoder or step 
on foot switch. 

 

Manual mode setting

 

1. Set the Operating flow to 0 in step 2 of the automatic cycle mode setting; 
2. Set the time to 0 in step 4 of the automatic cycle mode setting; 
At this point, the start and stop of the pump is controlled by the encoder or the foot 
switch. 
 
 

Summary of Contents for DIPump550

Page 1: ...DIPump550 intelligent peristaltic pump User Manual 2022 01 12 V1 4 Kamoer Fluid Tech Shanghai Co Ltd www kamoer com...

Page 2: ......

Page 3: ...ol RS485 communication control foot switch control External analog signal control 4 20mA 0 5V Power failure parameter memory support Working mode Speed mode Flow mode Way of working Fully automatic ci...

Page 4: ...ng the last three digits indicate the currently set working time When the first digit shows no beating the last three digits indicate the current remaining working time When the first digit is display...

Page 5: ...ristaltic pump when dialed to the left the pump rotates counterclockwise when it is dialed to the right the pump rotates clockwise The switch is ineffective during the operation of the peristaltic pum...

Page 6: ...cally 6 Press the encoder not long press the pump will run according to the set running time and stop time Note When the pump starts to cycle the digital tube displays the time countdown and follows 1...

Page 7: ...nd the running flow and stop time will be saved automatically 6 Press the encoder not long press the pump will run according to the set Operating flow and stop time 7 Press the encoder not long press...

Page 8: ...igits of 1 8 correspond to 8 different functions Select the function to be set and press the rotary encoder to enter the detailed setting interface of the function Specific function settings Note In t...

Page 9: ...here is no need to set this item Interface Setting Introduce Low level mode When the direction button is turned counterclockwise but the pump actually rotates clockwise select 0 to adjust the pump to...

Page 10: ...rform pump head flow calibration After entering the setting interface the calibration speed and calibration time will switch back and forth every 2 seconds the first digit shows 4 the middle one space...

Page 11: ...g interface and you can set the analog selection The last digit represents the analog option Interface Setting Introduce 0 5V 4 20mA 485 mode C007 Suction back function In the function selection inter...

Page 12: ...g interface and the drive enable setting can be performed The last digit represents the enable option Interface Setting Introduce Low level enable High level enable C009 Maximum speed setting This fun...

Page 13: ...gnal control mode Control the board by analog signal mode 0 5V 4 20mA Please connect the signal line according to the actual control mode Pin 1 reserved Pin 2 485 B Pin 3 485 A Pin 4 Foot switch Pin 5...

Page 14: ...452 Note the maximum flow in the above chart is obtained by testing water at 400 rpm with a new pump pipe aged for 30 minutes at room temperature about 25 for reference only Ambient temperature mater...

Page 15: ......

Page 16: ...Kamoer Fluid Tech Shanghai Co Ltd Add 4th Building No 49 Xiangjing Road Songjiang District Shanghai China Phone 86 021 67742578 Fax 86 021 67741776 Website www kamoer com...

Reviews: