Working mode - speed mode
Set the working mode to speed mode through the function menu, and the factory default is speed
mode. The parameters set in speed mode include speed (unit is rev/min), running time (unit is
second), stop time (unit is second).
Fully automatic cycle mode setting
1. When the pump is stopped, the working interface displays the current speed (rev/min);
2. Press and hold the encoder to enter the pump running time setting interface. At this time, the
first digit of the digital tube only displays the upper half ; rotate the encoder to set the running
time, and the setting range is 0 to 999 seconds;
3. Short press the encoder to switch to the pump stop time setting interface. At this time, the first
digit of the digital tube only displays the lower half ;
4. Rotate the encoder to set the stop time, the setting range is 0-999 seconds;
5. Long press the encoder to return to the speed display interface, and the running time and stop
time will be saved automatically;
6. Short press the encoder, the pump will start to cycle according to the set running time and stop
time;
7. Short press the encoder, the pump stops, and the digital tube switches to the speed display
interface.
Semi-automatic cycle mode setting
Just set the pump stop time to 0 in step 4 of the fully automatic circulation mode setting. At this time,
after the pump stops running according to the set running time, it will not start automatically again;
if you want to start again, short press the encoder.
Manual mode setting
In the fully automatic cycle mode setting, the second step sets the pump running time to 0, and the
fourth step sets the pump stop time to 0. At this time, the start and stop of the pump are controlled
by an encoder or a foot switch, and the encoder can be rotated to adjust the speed during
operation.