Kamoer Fluid Tech(Shanghai) Co., Ltd. DIP-KK300 V2 User Manual
10
Function
Display
Setting Description
0 means trigger mode (press the foot switch
once, the pump starts to work, press it again,
the pump stops working)
1 means low level trigger (long press the foot
switch, the pump works, release the foot
switch, the pump stops working)
2 stands for high level trigger (release the foot
switch, the pump works, long press the foot
switch, the pump stops working)
"C004" flow calibration
1)
On the function selection interface, rotate the encoder, when the corresponding function of
"C004" is displayed as "Flow Calibration", short press the rotary encoder to enter the setting
interface, and the flow of the pump head can be calibrated.
2)
After entering the setting interface, the first digit of the digital tube displays 4, a space in the
middle, and the last 2 digits represent time seconds.
3)
When the pump is stopped, it will switch to display "speed (RPM)" and "4 time" every 2
seconds. When the speed is displayed, the rotary encoder can be rotated to change the speed.
When the time is displayed, the rotary encoder can be rotated to change the time.
4)
After completing the setting, short press the rotary encoder to start the flow calibration work,
and the display interface will display "4 countdown time" during operation.
5)
After the countdown is over, the screen displays the volume input interface. You can input the
actual volume by rotating the rotary encoder. After completing the input, you can long press
the rotary encoder to save and return to the working interface.
Short press the rotary encoder to return to the function selection interface; long press the rotary
encoder (the setting takes effect) and return to the working interface.
Function
Display
Setting Description
set calibration speed
set calibration time
enter actual volume