7
KFMANUAL 0317
KAM CONTROLS, INC.
GLASSWARE ASSEMBLY
Remove glassware (Titration Cell Unit) from packaging or base.
Glassware should be assembled and disassembled separately
prior to installation on the unit base. This prevents the possibility
of spilling reagents on the base electronics.
If assembling a new unit, remove the Silica Gel from
packaging and pour into the Drying Tube
(A)
, or verify that the
existing Silica Gel inside the Drying Tube is blue. If not, replace.
Insert Drying Tube Assembly into the smallest opening of the
Generator Solution Cell lid
(B)
.
Unscrew Injection Port
(D and E)
and place a new rubber Sep-
tum
(C)
in the base of the Injection Port
(D)
. Screw the top of
the Injection Port
(E)
into the base and place the Injection Port
into the indicated opening on the Generator Solution Cell lid
(B)
.
Apply a thin layer of Special Grease to the neck of Detector
Electrode
(F)
and place in the remaining small opening of the
Generator Solution Cell lid
(B)
. Do not apply pressure to the
Electrode. Rather, let it seat under its own weight. Rotate it half
a turn backward and forward to ensure the grease is spread
evenly and a seal has been created.
Drop the ceramic Stir Bar
(J)
into the Generator Solution Cell.
Apply a thin layer of Special Grease to the neck of
Cathode Solution Cell
(G)
and its Stopper
(H)
. Set aside.
Pour entire contents of KAM Generator Solution A (100 ml)
into the Generator Solution Cell through the remaining hole
(should be the large hole near the center of the lid). The
Generator Solution A (100 ml) is pre-measured to the
appropriate volume.
Place the Cathode Solution Cell
(G)
into in the Generator
Solution Cell lid
(B)
. Do not apply pressure. It will seat under its
own weight. Rotate it half a turn backward and forward to
ensure the grease is distributed evenly around the circumfer-
ence of the seal. The sooner the hole is covered, the less mois-
ture will be absorbed by the chemicals.
1.
2.
3.
4.
5.
6.
7.
8.
4. OPERATION
B
J
C
D
E
G
H
F
A
NOTE: Starter Kit, including reagents, consumables and accessories, sold separately. See page 21
for details.
CAUTION: PROTECTIVE GEAR SHOULD BE WORN DURING ALL REAGENT HANDLING.