- 16 -
Connect solderless terminal to the output terminal
on the rear panel to output the comparator result.
NOTE :
See the connecting diagrams in fig. 12 when
using the relay output.
NOTE :
Do not apply the loading that might exceed
the relay specification.
6-4. Comparator Relay Output
2. Dimensions of Recommended Solderless Terminal
Relay specification
Loading voltage : 250V DC, 250V AC MAX.
ON resistance : 35
Ω
MAX.
Continuous loading current : 120mA MAX.
1. Connecting Diagram
①
AC/DC Loading
②
DC Loading (1)
+
+
−
VL
(AC/DC)
Load
+
+
−
+
−
VL (DC)
③
DC Loading (2)
+
+
−
+
+
−
VL (DC)
VL (DC)
Load
Load
Load
NOTE :
Sum of continuous load current
must not exceed the maximum
rating.
①
Circle-shaped terminal
②
Y-shaped (tip open type) terminal
●
When taking the comparator measurement with the input terminals are opened, such as when
the resistance measurement, comparator always shows "Hi" result with "OL" sign on LCD. To
prevent this problem, add GO/STOP signal to your inspection system in reference to the circuit
diagram shown in Fig. 11.
※
Reference circuit diagram
(in case of DC)
+
+
−
+
−
VL (DC)
Loading
Contact point like magnet relay
Fig. 11
Fig. 12
Fig. 13
Important Note for Comparator Measurement
Your inspection
system
GO/STOP signal from your inspection system
※
Run the comparator when it is ON.