13

IX. Maintenance (Notice)

9.1 Please wipe the electrodes with clean wet cloth before use after long time place. Then
immerse into the water 3-5 hours;
9.2 Please keep the sensor clean, avoid mud and grease;
9.3 Please keep the electromagnetic velocity display clean and dry;
9.4 Please remember to change the battery.(Battery is 1.5V*5, long time place should
remove the internal battery).

X. Fault handling

NO.

Alarming

Information

Treatment

1

Low press

alarm

Please replace battery

2

Exciting alarm

1. Excitation line loose;
2. Excitation line fault;
3. Return to factory.

3

Bare electrode

1. Velocity measuring height
setting is not current;
2. Water level sensor damage;
3. Return to factory.

XI. Product complete set

10.1 MGG/KL-DCB Portable velocity display

1set

10.2 Velocity sensor

1set

10.3 Velocity meter tail

1set

10.4 Steel measure tape

1pcs

10.5 Allen wrench

3pcs

10.6 Measuring insert rod(Order length)

1set

10.7 Measuring steel wire(Order length)

1pcs

XII. Random documents

11.1 Operation instruction

1pcs

11.2 Product certificate

1pcs

11.3 Package

1pcs

11.4 Feedback

1pcs

Summary of Contents for MGG/KL-DCB

Page 1: ...MGG KL DCB Portable Electromagnetic Velocity Meter Operation Manual...

Page 2: ...measurement multiple parameters display self diagnosis alarming RS485 optional and storage function MGG KL DCB is the series model of our portable electromagnetic velocity flow meter This series incl...

Page 3: ...parameter setting 5 7 1 Velocity meter model menu 6 7 2 Engineering mode menu 7 Instrument usage 8 8 1 Velocity measuring 8 8 2 Flow measuring 9 8 3 Measurement requirements 10 8 4 Operating mode 11 8...

Page 4: ...ter conservancy II Main feature No movable parts no winding No choked flow no maintenance avoid troubles of other calibration and testing Micro power design can last 80 hours when replace the battery...

Page 5: ...So we install a pair of electrodes on the pipe wall which orthogonal with sounding pipe axis and magnetic line The electrodes can detect the induced electromotive force It should be directly proporti...

Page 6: ...ted coupler plug Please directly insert the plug and tighten the screws 6 1 2 Display interface Model Model display interface Model display interface 6 1 3 Battery and communication port The battery i...

Page 7: ...4 6 2 Velocity sensor The velocity sensor consists of velocity sensor and tail cable Velocity Sensor The Tail...

Page 8: ...assword input Under Engineering Mode The password must be inputted when modify status and parameter Password can be set arbitrarily in 5 digit number between 00000 99999 the default is 00000 In the wo...

Page 9: ...s 1 m h Velocity digit 0 3 0 0 000 Time 1 10s Time 1 10s Interval 1 10s Interval 1 10s Value number 0 999 Reverse adjustment 0 off 1 on Record 0 999 Auto shutdown 0 off 1 on Auto shutdown 0 on 1 off...

Page 10: ...uiry of value through page turning to judge value property Auto shutdown 1 is Time limited model 1800s auto shutdown 60s auto shutdown without any operation 0 is continuous working model Backlight 0 i...

Page 11: ...rd It is adjustable parameter the factory default is 0000 Velocity cutoff Low velocity cutoff display unit is m s Model Transmission address The range of mailing address is 1 255 default is 1 Model Tr...

Page 12: ...uring margin velocity measuring height side slope ratio channel bottom width water depth water level adjustment Model water level can be automatically collected by depth sensor Note The measuring meth...

Page 13: ...0 2 V0 8 2 3 poin t 0 2 0 6 0 8h Vm 0 25 V0 2 2 V0 6 V0 8 4 poin t 0 2 0 4 0 7 0 9h Vm 0 25 V0 2 V0 4 V0 7 V0 9 5 poin t Close to surface 0 2 0 6 0 8h Riverbed Vm 0 1 V0 3V0 2 3V0 4 2V0 8 VS 6 poin t...

Page 14: ...r parallels to the flow direction Waiting a moment for stable running of sensor then resend the data Avoid the primary interference signal input and eliminate the electrode oxide film interference 8 4...

Page 15: ...calibration both medium and sensor 8 6 Historical data record and query Model display can record 1000 items historical data 8 6 1 Data record and storage The instrument will record internal storage nu...

Page 16: ...1 Low press alarm Please replace battery 2 Exciting alarm 1 Excitation line loose 2 Excitation line fault 3 Return to factory 3 Bare electrode 1 Velocity measuring height setting is not current 2 Wat...

Page 17: ...Dear Customer Thank you for your business Please complete the form below to improve our quality and service Your cooperation will be highly appreciate Product Model No Inspector Certification Problem...

Reviews: