K203/K205ULT Operating Manual
Product Description
5
Page 1: ...parameters may damage the unit or void your warranty Please contact K2 before attempting to change advanced parameters Unfamiliar with the operation of a controller Use the video tutorials on our web...
Page 2: ...adjustments and maintenance must be performed by qualified service personnel Do not use this appliance for other than its intended use Do not cover the front grille or block the rear air entry by plac...
Page 3: ...ay to prevent any overflow Ventilation All models must have clear and unobstructed ventilation from the entire surface area of the front grille Temperature Controller 3 SET RST Enter a parameter setti...
Page 4: ...ressor stop High temperature compresso time delay Low temperature compressor work in non forced refrigeration mode Low temperature compressor is closed in non forced refrigeration mode Forced refriger...
Page 5: ...logger 15 0 15 0 0 C9 Condenser probe calibration High temp compressor loop 15 0 15 0 0 C10 Ambient temp probe calibration 15 0 15 0 0 C11 High temp evaporator probe calibration 15 0 15 0 0 C13 The Mi...
Page 6: ...K203 K205ULT Operating Manual Product Description 5...
Page 7: ...K203 K205ULT Operating Manual Maintenance 6 Alarm Code...
Page 8: ...ntact us via our chat feature at www k2sci com 2 Email support k2sci com 3 Call 800 218 7613 When storing vaccines you may be required to preform a field validation test A NIST callibrated external da...
Page 9: ...f cabinet C13 50 0 40 C15 Maximum temp of high temp evaporator when low temp compressor turned off C7 10 0 12 H1 The cycle time of capillary heating wire 1 300 108 H2 The working time of capillary hea...