Instruction Manual - G85
20
6.4 TUNE
The TUNE option is for
calibration
of the
displayed temperature.
If there is an offset between the value on the display and any measurements made with a
high precision external temperature sensor, this can be corrected.
The new temperature reading will be kept as the displayed value, with temperature control
conducted on this basis.
Follow these steps to navigate in the TUNE option.
Action
Key
Display
Press and hold UP+DOWN key for
3 seconds to enter the menu.
The menu "USER" will appear in
the display.
Press DOWN key until the option
“TUNE” appears in the display.
Press and hold setpoint key to
change values.
While holding the setpoint key in,
press either the arrow UP or arrow
DOWN key to select the value
temperature.
When the required setting is
activated, let go of the setpoint
key.
Press and hold UP+DOWN key for
3 seconds to exit the menu.
Any change in the factory set calibration should always be based on very
certain
temperature measurements. K-SYSTEMS Kivex Biotec only recommend the
use of a high quality PT-100 sensor that is calibrated with the precision instrument
used for measuring according to the manufacturers specifications. Also ensure an
optimal contact between the sensor and the place measured. Use only a sensor type
that is correct for the purpose and of correct size. If in doubt contact your local
distributor or K-SYSTEMS Kivex Biotec.