2.2 Mounting the sensor
The sensor mounting location should be chosen with care to ensure reliable readings
from the process. Some basic rules, described in this section, apply to all sensor mod-
els. The model specific instructions can be found in Chapter 9, “Sensor specifications”.
For the Sanitary compact refractometer PR-23-AC see Section 9.3, for the Probe san-
itary refractometer PR-23-AP see Section 9.4, for the Process probe refractometer
PR-23-GP see Section 9.6, for the Teflon body refractometer PR-23-M and Teflon body
semicon refractometer PR-23-MS see Section 9.8 and for the Saunders body refrac-
tometer PR-23-W see Section 9.9. For mounting an ATEX/FM/CSA approved sensor
in explosive atmosphere, see Section 9.10.2. For mounting an intrinsically safe refrac-
tometer PR-23-...-IA, see Section 9.11. For mounting of the Safe-Drive™ system with
the PR-23-SD sensor, see Section 11.4.
2.2.1 Choosing sensor mounting location
A K-Patents inline refractometer sensor can be located either indoors or outdoors in
most climates. However, when a sensor is located outdoors, some basic protection
against direct exposure to sunlight and rain should be provided. Special care should be
taken if the pipe wall is translucent (e.g. of fiberglass), as light from outside reaching
the prism through the pipe wall may disturb the measurement.
The mounting location needs to be such that sediments or gas bubbles cannot accu-
mulate by the sensor. Good flow velocity is essential in keeping the prism clean.
Important:
If the process pipe vibrates, support the pipe. A vibrating pipe might
damage the in-line sensor mounted on it.
Always check that the sensor head is kept cool enough; the sensor head should not be
too hot to keep a hand on. The sensor cover should not be exposed to high temperature
radiation. In most cases, draft and natural convection provide sufficient air cooling if
the air gets to flow freely around the sensor head.
Additional cooling is necessary when the ambient temperature is higher than 45 °C
(113 °F) or when the process temperature is above 110 °C (230 °F) and the ambient
temperature is above 35 °C (95 °F). The air cooling is improved by blowing pressurized
air against the sensor cover. The pressurized air can be supplied by the ventilation
system. If no air is available it is also possible to wind a copper coil for cooling water
around the sensor head cover.
Important:
Always mount the sensor so that the interconnecting cable points down-
wards from the sensor head.
Summary of Contents for PR-23-AC
Page 1: ...PROCESS INSTRUMENTS PROCESS REFRACTOMETER PR 23 INSTRUCTION MANUAL IM EN PR23 Rev 1 88 ...
Page 2: ......
Page 4: ...PR 23 instruction manual ...
Page 17: ...2 Inline refractometer sensor 7 2 2 2 PR 23 mounting guide ...
Page 26: ...16 PR 23 instruction manual ...
Page 29: ...4 Prism wash systems 19 Figure 4 1 A prism wash system for steam non sanitary ...
Page 30: ...20 PR 23 instruction manual Figure 4 2 A sanitary prism wash system for steam ...
Page 31: ...4 Prism wash systems 21 Figure 4 3 Wiring for a prism wash system for steam ...
Page 33: ...4 Prism wash systems 23 Figure 4 5 A sanitary prism wash system for high pressure water ...
Page 34: ...24 PR 23 instruction manual Figure 4 6 Wiring for a prism wash system for high pressure water ...
Page 84: ...74 PR 23 instruction manual ...
Page 117: ...9 Sensor specifications 107 9 7 5 PR 23 RP dimensions Figure 9 22 PR 23 RP 73 M20 ...
Page 160: ...150 PR 23 instruction manual 11 4 2 Wiring Figure 11 9 PR 23 SD system wiring ...
Page 174: ...164 PR 23 instruction manual ...
Page 186: ...176 PR 23 instruction manual ...
Page 194: ...184 PR 23 instruction manual ...
Page 198: ...188 PR 23 instruction manual ...
Page 200: ...190 ...
Page 208: ...198 ...
Page 210: ...200 ...
Page 211: ...201 E DTR command selection tree ...
Page 212: ...202 ...
Page 213: ...203 F STR Divert mode command selection tree ...
Page 214: ...204 ...