JYE Tech                                                                - 4 -                                                                  www.jyetech.com

   

(a) Under compensation 

(b) Right compensation 

(c) Over compensation 

Fig 4 

Fig 5 

Fig 6 

3)

 

Set vertical sensitivity to 2V (V.Sen 1 switch in position 

1V

 And V. Sen 2 switch in position 

X2

). 

4)

 

Set the coupling switch to the 

DC

 position. (See Fig 3)   

5)

 

Press the [ V.POS ] button, and adjust 0V indicator to the position shown in Fig. 3. 

6)

 

Press the [ SEC/DIV ] button, and set the  Time-base to 0.5ms. Waveform shown in Fig. 3 should be 

seen.  (If  you  find  the  display  is  unstable,  you  can  press  [LEVEL]  and  turn  the  knob  so  that  a  stable 

display is obtained.) 

7)

 

Changing position of V. Sen 2 switch you should see that the waveform amplitude changes accordingly. 

You can compare the amplitude readings. 

8)

 

Changing  the  Time-base  setting  to  0.2ms,  for  instance,  you  should  see  the  signal  waveform  widens 

accordingly. Compare readings and experiment with other settings. 

9)

 

Now, select 

AC

 with the coupling switch and you should see that the waveform is shifted down to a 

position where the VPos indicator is at its midway point.   

 

Example 2 Observe saw signal (Learn how to use trigger modes) 

The schematic on the right is a simple sawtooth signal generator. 

We are going to use the scope to observe its output. Build the circuit 

according  to  the  schematic.  Connect  power  and  oscilloscope  as 

indicated. 

1)

 

Select coupling switch to DC. Set V. Sen. 1 switch to the 

1V

  position  and  V.  Sen.  2  switch  to  the 

X2

  position. 

(i.e.  sensitivity  is  set  to  2V/Div).  Adjust  the  0V  to  the 

second last point on the vertical scale. (See Fig 4) 

2)

 

Power  the  circuit.  Select  Time-base  to  0.5ms  (or  other 

depending  on  type  of  transistor  used).  You  should  see  a 

waveform similar to that shown in Fig 4.   

3)

 

Select  AUTO  trigger  mode  and  change  the  trigger  level. 

You should be able to observe that the waveform stabilizes 

itself  when  it  intersects  with  trigger  level  (the  internal 

trigger  source  should  be  selected  here).  Otherwise,  the 

waveform  will jump and you won

t be able to get a clear 

view of it.   

4)

 

Change  the  trigger  mode  to  NORM,  and  vary  the  trigger 

level.  Notice  that  the  waveform  only  update  when  they 
intersect  with  the  trigger  level  (internal  trigger  source 

should also be selected here). Otherwise, it will stall.   

5)

 

Press the [HOLD] button to put the oscilloscope into the HOLD state. 

HOLD

 is displayed. Now you 

can shift the waveform back and forth to view the un-displayed waveform. Pressing [HOLD] again will 

release the HOLD state and put oscilloscope back into captured signal sample state.   

Use 10X Probe 

10X  probe  offers  the  advantage  of  higher  input  voltage  range,  higher 

impedance,  and  less  interference  to  the  circuits  being  measured.  In  order  to 
effectively use 10X probe the compensation cap trimmer (see Fig 5) must be 

calibrated. This can be done using the built-in 1KHz/5V test signal. 

Set  the  probe  attenuation  to  1/10  by  setting  the  attenuation  switch  in  the 

probe  handle  to  the 

10X

  position.  Insert  the  probe  tip  into  the  centre 

conductor of M.F.T. connector. Set oscilloscope Time-base to 0.2ms/Div and 

adjust trigger level so that a stable waveform display is obtained. Tune the cap 

trimmer carefully (see Fig 5) so that right compensation figure is reached 

(refer to Fig 6). 

At using 10X probe remember to multiply vertical reading by 10 for true results. If, for example, the displayed 

signal amplitude is 2.5V the real signal amplitude is actually 25V. 

 

Summary of Contents for DSO 082

Page 1: ...d respectively below Screen 1 Vertical position indicator the small triangle on the left border of the grid indicates the 0 volt input level Its position can be changed up or down by pressing VPOS and...

Page 2: ...h blocks the DC component of input signal When this switch is set to GND the input signal is disconnected and oscilloscope input is connected to ground Vertical Sensitivity Select Switches There are t...

Page 3: ...of oscilloscope It serves two purposes 1 Test Signal output 2 External Trigger input The function of M F T is selected by LEVEL key and indicated by the small letter at the right of trigger mode indic...

Page 4: ...ther depending on type of transistor used You should see a waveform similar to that shown in Fig 4 3 Select AUTO trigger mode and change the trigger level You should be able to observe that the wavefo...

Page 5: ...ry power Power Off To power off the oscilloscope push down the adjustment knob and hold it for about 3 seconds The oscilloscope will enter stand by mode where most of circuits are powered off Only key...

Page 6: ...d in on the screen 5 Press the LEVEL button then HOLD Scope screen will be sent to PC as a bitmap file Using the External Trigger 1 To Select the trigger source Press LEVEL to highlight the trigger le...

Page 7: ...put Voltage 50Vpp for 1X probe Couple Range DC AC GND Horizontal Max Real time Sample Rate 50MSPS Time base Range 0 2us Div 10minute Div Record Length 2000 16000 points user selectable in steps of 200...

Page 8: ...refurbished parts ii replace the oscilloscope with a new or refurbished one or iii refund to you all or part of the purchase price of the oscilloscope All replaced parts and oscilloscopes for which a...

Reviews: