JYE DSO 082 Manual Download Page 3

                                                                                                                      DN082-02v02 

JYE Tech                                                                - 3 -                                                                  www.jyetech.com

   

LEVEL 

Selects trigger level

. When this button is pressed, the trigger level adjustment indicator will be highlighted 

and the trigger level can be adjusted by turning the adjustment knob. 

Pressing this button again will toggle the 

trigger source between internal and external.

   

HOLD 

Toggle the oscilloscope between HOLD and RUN states.   
 

4.

 

Connectors and Terminals 

Power Supply Connector 

This connector is located at bottom right of the panel. A 9 

 12V DC power supply with minimum 500mA 

capacity should be used. Even though there is over-voltage protection, a power supply  voltage higher than 13V 

should be avoided. The centre conductor of the connector is the positive pole of the power supply.   

Signal Input Connector 

This connector locates at bottom left of the panel. This is a standard BNC connector for oscilloscope probe. 

Please note that with a 1X probe the maximum peak-to-peak voltage applied to this input should not be greater 

than 50V. If measurement for larger voltage is needed appropriate divider (use 10X probe, for example) should be 

provided. 

USB Connector 

This connector locates at the right side of oscilloscope. This is a 5-pin mini USB B type connector. It is for 

firmware upgrading and data transferring to/from PC. 

M.F.T. Connector 

The Multi-Function Terminal (M.F.T.) locates at the left side of oscilloscope. It serves two purposes: 

1)

 

Test Signal output 

2)

 

External Trigger input 

The function of M.F.T. is selected by [LEVEL] key and indicated by the small letter at the right of trigger 

mode indicator. The table below summarizes the functions of M.F.T. 

Indicator  Trigger Source 

Test Signal Output 

 i 

 

Internal 

Enabled 

 e 

 

External 

Disabled 

 

5.

 

Basic Operations 

Power Up 

Connect power adapter to the connector at bottom-right and push the [KNOB] button. The oscilloscope first 

enters bootloader and stays for 1 

 2 seconds, and waits for firmware upgrade request. If no upgrade request is 

detected control is passed to main firmware. Logo and firmware version information will be briefly displayed and 

then oscilloscope enters normal working state. 

Set Parameters 

The basic and most frequently used oscilloscope parameters are: 
1)

 

Vertical related 

 sensitivity, vertical position [VPOS], and AC/DC coupling.   

2)

 

Horizontal related 

 Time-base [SEC/DIV] and horizontal position [HPOS]. 

3)

 

Trigger 

 trigger modes [MODE], edges [SLOPE], source, and level [LEVEL]. 

Please refer to the descriptions in Section 4 for how to use these controls to change the parameters. 
 

Example 1 Observe the test signal (Practice basic operations) 

1)

 

Place  the  tip  of  the  probe  on  the  centre  conductor  of  the  BNC 
connecter  marked  M.F.T.  This  is  the  output  of  built-in 

1KHz/5Vpp  test  signal. 

Remember  to  set  the  attenuation 

switch at probe handle to 1X position. 

2)

 

Press  the  [LEVEL]  button  to  set  the  trigger  source  to  internal 

(letter 

i

  should  be  displayed.  Otherwise  you  won

t  get  test 

signal output.) 

Fig 3 

Summary of Contents for DSO 082

Page 1: ...d respectively below Screen 1 Vertical position indicator the small triangle on the left border of the grid indicates the 0 volt input level Its position can be changed up or down by pressing VPOS and...

Page 2: ...h blocks the DC component of input signal When this switch is set to GND the input signal is disconnected and oscilloscope input is connected to ground Vertical Sensitivity Select Switches There are t...

Page 3: ...of oscilloscope It serves two purposes 1 Test Signal output 2 External Trigger input The function of M F T is selected by LEVEL key and indicated by the small letter at the right of trigger mode indic...

Page 4: ...ther depending on type of transistor used You should see a waveform similar to that shown in Fig 4 3 Select AUTO trigger mode and change the trigger level You should be able to observe that the wavefo...

Page 5: ...ry power Power Off To power off the oscilloscope push down the adjustment knob and hold it for about 3 seconds The oscilloscope will enter stand by mode where most of circuits are powered off Only key...

Page 6: ...d in on the screen 5 Press the LEVEL button then HOLD Scope screen will be sent to PC as a bitmap file Using the External Trigger 1 To Select the trigger source Press LEVEL to highlight the trigger le...

Page 7: ...put Voltage 50Vpp for 1X probe Couple Range DC AC GND Horizontal Max Real time Sample Rate 50MSPS Time base Range 0 2us Div 10minute Div Record Length 2000 16000 points user selectable in steps of 200...

Page 8: ...refurbished parts ii replace the oscilloscope with a new or refurbished one or iii refund to you all or part of the purchase price of the oscilloscope All replaced parts and oscilloscopes for which a...