3 Preparing the measuring probe
22
4. Unscrew the sealing screw M12 × 1.5, replace the cable fitting (D) with O-ring 11 × 1.5 mm, and
tighten on the measuring probe (max. 1 Nm).
For the initial startup, the cable fitting with O-ring is in the accessories set and not fitted to the mea-
suring probe.
5. Guide the external voltage supply connection out through the cable fitting (D) and connect to the
spring-cage terminal (C). Make the sure the polarity (+ and -) is correct!
6. Manually tighten the union nut of the cable fitting (D). Place the battery compartment cover (B) on
the measuring probe. Make sure that the external voltage supply connection is not damaged by
pinching or bending!
7. Tighten the screw (A) clockwise on the battery compartment cover (B) (max. 1 Nm).
8. Fit the antenna (E), see chapter 3.4.1 "Installing the antenna", Page 23.
For the initial startup, the antenna is in the accessories set and not fitted to the measuring probe.