12 IO-Link data
60
TC curve
Designation
Index Sub-
index
Data type
Value range
Default Ac-
cess
right
a
a
RW = Read and write access
RO = Read access only
WO = Write access only
Description
Temperature[0]
170 to
173
for MR
1 to 4
1
TFLOAT
-20 to 150 °C
0
RW
Temperature co-
efficient[0]
2
TFLOAT
0 to 8
0
RW
Temperature[1]
3
TFLOAT
-20 to 150 °C
0
RW
Temperature co-
efficient[1]
4
TFLOAT
0 to 8
0
RW
Temperature[2]
5
TFLOAT
-20 to 150 °C
0
RW
Temperature co-
efficient[2]
6
TFLOAT
0 to 8
0
RW
Temperature[3]
7
TFLOAT
-20 to 150 °C
0
RW
Temperature co-
efficient[3]
8
TFLOAT
0 to 8
0
RW
Temperature[4]
9
TFLOAT
-20 to 150 °C
0
RW
Temperature co-
efficient[4]
10
TFLOAT
0 to 8
0
RW
Temperature[5]
11
TFLOAT
-20 to 150 °C
0
RW
Temperature co-
efficient[5]
12
TFLOAT
0 to 8
0
RW
Temperature input
Designation
Index Sub-
index
Data type
Value range
Default Ac-
cess
right
a
a
RW = Read and write access
RO = Read access only
WO = Write access only
Description
Filter time con-
stant
180
0
TFLOAT
0 to 25
2
RW
Offset
181
0
TFLOAT
-10 to+10 °C
0
RW
Connection type
182
0
TENUM
TENUM values:
0x00: 2-wire
0x01: 3-wire
0x00
RW
Temperature in-
put activation
183
0
TENUM
TENUM values:
0x00: Inactive
0x01: Active
0x01
RW
Calibration timer
Designation
Index Sub-
index
Data type
Value range
Default Ac-
cess
right
a
Description
Calibration inter-
val
280
0
TUINT16
0 to 9999 d
0
RW
d = days
Summary of Contents for digiLine CR HT10
Page 2: ......
Page 6: ...Contents...
Page 16: ...3 Device Description 16...
Page 26: ...4 Identifying the device version 26...
Page 50: ...10 Troubleshooting in the event of faults 50...
Page 70: ...12 IO Link data 70...
Page 78: ...14 Operation maintenance and care 78...
Page 83: ......