35
6 Technical data (extract from data sheet)
Tran
s
ducer
s
hort circuit/brea
k
Binary input
s
/output
s
(option)
Output
s
Interface
s
S
hort-circuit
1
Brea
k
1
Thermocouple
not detected
detected
Resistance thermometer
detected
detected
Resistance transmitter
not detected
detected
Potentiometer
not detected
detected
Voltage
≤
± 210mV
not detected
detected
Voltage > ± 210mV
not detected
not detected
Current
not detected
not detected
1
Programmable reaction of device, e.g. triggering alarm
Input or output
configurable as input or output
Number
8, 16 or 24, depending on the device version,
to DIN VDE 0411, Part 500; max. 25Hz, max. 32V
Input
- level
logic "0": -3 to +5V (input current max. ±1mA),
logic "1": 12 to 30V (2.5mA
≤
input current
≤
5mA)
- counting frequency
8Hz
High-speed input
the first two binary inputs of each module (B1, B2, B9, B10, B17, B18),
if the module is not fitted with relays or 6 analog inputs
- task
count function, e.g. for flow measurement
- counting frequency
10kHz
Output
- type
open-collector output, switches relative to positive voltage
- level
logic "0": transistor is inhibited
(max. permissible voltage across switching transistor 30V, max. leakage current 0.1 mA)
logic "1": transistor is switched on
(max. voltage across switching transistor 1.6V, max. current 50 mA)
- sampling cycle
at least 1sec (1Hz)
1 relay (ex-factory)
changeover (SPDT), 3A, 230V AC
1
6 relays (option)
changeover (SPDT), 3A, 230V AC
1,2
1
With resistive load.
2
It is not permissible to mix SELV circuits and supply circuits.
RS232/RS485 (connector 7)
- protocol
- baud rate
- modem
- connector
- external inputs
Qty. 1, switchable between RS232 and RS485
Modbus master, Modbus slave and barcode reader
9600, 19200, 38400
can be connected
SUB-D
via the Modbus master/slave function, 24 analog and 24 binary
RS232 for barcode reader (connector 2)
- protocol
- baud rate
- connector
- external inputs
Qty. 1
Modbus master, Modbus slave and barcode reader
9600, 19200, 38400
SUB-D
via the Modbus master/slave function, 24 analog and 24 binary
Ethernet (connector 6)
- quantity
- protocols
- baud rate
- connector
- data format
max. 1
TCP, IP, HTTP, DHCP, SMTP, ModbusTCP
10Mbits/sec, 100Mbits/sec
RJ45
HTML
USB host (connector 5)
- quantity
- use
- max. current
2 (or 1 with stainless steel front),
connector 5 and front connector (not with stainless steel front); no parallel operation
for connecting a memory stick
100mA
USB device (connector 15)
- quantity
- use
2 (or 1 with stainless steel front),
connector 15 and front connector (not with stainless steel front); no parallel operation)
for connecting to the (master) computer
Summary of Contents for 706581/18-321-33/020,444
Page 2: ......
Page 4: ...Contents 9 Index 43 ...
Page 10: ...1 Introduction 10 ...
Page 14: ...2 Identifying the instrument version 14 ...
Page 30: ...4 Electrical connection 30 ...
Page 32: ...5 Functional test 32 ...
Page 37: ...37 7 Declaration of Conformity ...
Page 38: ...7 Declaration of Conformity 38 ...
Page 39: ...39 8 Type Examination Certificate ...
Page 40: ...8 Type Examination Certificate 40 ...
Page 41: ...41 8 Type Examination Certificate ...
Page 42: ...8 Type Examination Certificate 42 ...
Page 45: ......
Page 46: ......
Page 47: ......