Parameter "rES pH": resolution of the pH display
The parameter "rES pH" can only be accessed when the "pH" measurand has
been selected.
✱
Use the
and
keys to set the resolution:
✱
Press the
key to save the setting and move to the next parameter or
✱
press the
key to save the setting and move to the next submenu of the
setting menu or
✱
press the
key to save the setting, quit the device settings, and return to
measurement mode.
Parameter "CAL": selection of the number of calibration points
The parameter "CAL" can only be accessed when the "pH" measurand has
been selected.
✱
Use the
and
keys to set the number of calibration points:
✱
Press the
key to save the setting and move to the next parameter or
✱
press the
key to save the setting and move to the next submenu of the
setting menu or
✱
press the
key to save the setting, quit the device settings, and return to
measurement mode.
Menu/parameter
Setting option
Meaning
0.1
Resolution in tenths pH
0.01
Resolution in hundredths pH
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
Menu/parameter
Setting option
Meaning
1-Pt
1 point calibration
(only offset, slope -59.2 mV/pH)
2-Pt
2 point calibration
(n one further buffer)
3-Pt
3 point calibration
(n one one alkaline buffer)
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL