8.2 Storing the sensor in the storage container
✱
Slide the union nut (2) onto the sensor shaft.
✱
Slide the O-ring of the union nut (2) onto the
sensor shaft.
✱
Insert the sensor into the storage container
filled with
electrolyte
(to a max. level of 2
cm).
✱
Seal the storage container (3) with the union
nut (2).
8.3 Removing the electrode from the storage container
✱
Fully
unscrew the union nut (2). Then take the sensor out of the
storage container.
Attention:
It is essential to keep to the
procedure described below, as
otherwise pressure buildup will
destroy the sensor membrane!
The sensor must
not be stored
dry
!
Attention:
Do not move the sensor in the
wet-storage container once the
screw fitting is closed!
The positive or negative pressure
that this produces will destroy the
membrane.
(1)
(2)
(3)