Hold the sensor as shown in the above diagram.
✱
At the level of the vent opening (2), slide the vent wedge (3) under
the silicone tubing.
✱
Screw the membrane cap (4) all the way down
(excess electrolyte will escape).
✱
Remove the vent wedge (3).
✱
Use an absorbent cloth to soak up excess electrolyte.
The ammonia sensor is now ready for startup.
Attention:
The electrolyte must not get into the plug-type
head (1) of the sensor! This causes contact problems.