Remote control
Command
Parameter
Response of circulator
in_sp_02
none
Working temperature „t 3“
in_sp_03
none
High temperature warning limit „t High“
in_sp_04
none
Low temperature warning limit „t Low“
in_sp_07
none
Pump pressure stage
in_par_01
none
Te - Time constant of the external bath.
in_par_02
none
Si - Internal slope
in_par_03
none
Ti - Time constant of the internal bath.
in_par_06
none
Xp control parameter of the internal controller.
in_par_07
none
Tn control parameter of the internal controller.
in_par_08
none
Tv control parameter of the internal controller.
in_mode_01
none
Selected setpoint:
0 = Setpoint „t 1“
1 = Setpoint „t 2“
2 = Setpoint „t 3“
in_mode_05
none
Circulator in Stop/Start condition:
0 = Stop
1 = Start
12.4. Status messages
Status messages
Description
00 MANUAL STOP
Circulator in „OFF“ state.
01 MANUAL START
Circulator in keypad control mode.
02 REMOTE STOP
Circulator in „r OFF“ state.
03 REMOTE START
Circulator in remote control mode.
12.5. Error messages
ERROR MESSAGES
Description
-01 LOW LEVEL ALARM
Low liquid level alarm.
-02 REFRIGERATOR ALARM
Control cable of the refrigerated circulator or
MVS solenoid valve controller short-circuited
or interrupted.
-03 EXCESS TEMPERATURE WARNING High temperature warning.
-04 LOW TEMPERATURE WARNING
Low temperature warning.
-05 WORKING SENSOR ALARM
Working temperature sensor short-circuited or
interrupted.
46