![Julabo MA-12 Operating Manual Download Page 16](http://html1.mh-extra.com/html/julabo/ma-12/ma-12_operating-manual_2030709016.webp)
Preparations
5. Preparations
5.1. Installation
•
Place the circulator in an upright position.
•
Place the unit on an even surface on a pad made of
non-
flammable
material
5.2. Bath fluids
Caution:
Carefully read the safety data sheet of the bath fluid used, particularly with
regard to the fire point!
If a bath fluid with a fire point of
≤
65 °C is used, only supervised operation is
possible.
Water:
The quality of water depends on local conditions.
•
Due to the high concentration of lime, hard water is not suitable for
temperature control because it leads to calcification in the bath.
•
Ferrous water can cause corrosion - even on stainless steel.
•
Chloric water can cause pitting corrosion.
•
Distilled and deionized water is unsuitable. Their special properties cause
corrosion in the bath, even in stainless steel.
Recommended bath fluids:
Bath fluid
Temperature range
soft/decalcified water
5 °C to 80 °C
See website for list of recommended bath fluids.
ATTENTION:
The maximum permissible viscosity is 50 mm
2
/s
16