Operator responsibility – Safety instructions
8
Immersion cooler
FT902
Working temperature range
°C
-90 ... 30
Temperature stability
°C
±1
Temperature selection
digital
Temperature indication
LED
Resolution °C
0.1
Temperature control
PID1
Cooling capacity
°C
Medium ethanol
kW
+20 10 -20 -40 -80
0.27 0.27 0.24 0.2 0.07
Cooling compressor
2-stage
Refrigerant
230 V / 50 Hz
115 V / 60 Hz
R404A/R23
R404A/R23
Electrical connections:
Pt100 external sensor
Pt100
Overall dimensions (WxDxH)
cm
38x55x60
Immersion probe (Lxdia.)
cm
-------
Immersion probe, flexible (Lxdia.)
cm
65x1.5
Connection tubing (L)
cm
160
Noise level, distance 1 m
dBA
60
Weight kg
50
Ambient temperature
°C
5 ... 35
Mains power connection
V/ Hz
230 V/50 Hz
190-253 / 50
Current input (at 230 V)
A
5
Mains power connection
V/ Hz
115 V/60 Hz
90-127 / 60
Current input (at 115 V)
A
7
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.