![Julabo DYNEO DD Original Operating Manual Download Page 38](http://html1.mh-extra.com/html/julabo/dyneo-dd/dyneo-dd_original-operating-manual_2030727038.webp)
Product description
38
Technical data
DYNEO DD-601F
Working temperature range
°C
-35 … +200
Cooling capacity
°C
+20
+10
0
-10
-20
-30
R452A*
kW
0.60
0.54
0.50
0.33
0.19
0.07
R449A
kW
0.60
0.54
0.44
0.27
0.16
0.04
Refrigerants
R449A, R452A**
Dimensions
Dimensions (W x D x H)
cm
36 x 46 x 74
Usable bath opening
cm
22 x 15
Bath depth
cm
20
Volumes min. … max.
l
8.0 … 10.0
Weight with circulator
kg
38.2
Performance data
Mains connection
100 V
50/60 Hz
115 V
60 Hz
200-230 V*
50/60 Hz
Current consumption
(nominal/total)
A
11/15
7/12
3-4/16
Switzerland
A
3-4/10
Great Britain
A
3-4/13
* permissible voltage deviation ±5 %
** at 100 V 50/60 Hz