If the DPR300 pulser is operated in external trigger mode, then the Trig / Sync connector is used for
receiving a positive going 3V to 5 V external trigger pulse. Triggering of the pulser will occur
synchronously with the rising edge of the trigger pulse.
When triggering the DPR300 pulser from an external source, it is important to ensure that the pulse
repetition frequency does not exceed limits defined later in this text.
6.1.17
RS-232 Interface Connectors
These connectors are a pair of RJ45 receptacles through which computer control of the DPR300 is
affected on units with the remote PC control option installed. An RS-232 serial-interface port on the
control computer is connected to the Input RJ45 receptacle using the eight-conductor reversing RJ45
cable and the DB-9 to RJ45 adapter supplied with the DPR300. When control of other DPR300
instruments is desired, they may be added in a daisy-chain fashion by connecting a reversing RJ45
cable from the RS-232 Output connector on one instrument to the RS-232 Input connector of the
next instrument.
6.1.18
AC Input Receptacle
This receptacle is standard power receptacle with fuses. Any supply voltage from 100VAC to
240VAC at 50 Hz or 60 Hz may be applied.