X. Error Codes
During the initial power-on testing it is possible the scale may
show error message.
The meaning of the error messages is described below.
Error
codes
Possible causes
Remedy
E1
EPROM data lose.
Recalibrate the scale.
E2
Remote
platform
is
not
connected to the scale when
press LOCAL/REMOTE key to
change to remote platform.
Connect the remote
platform properly to
the scale.
E3
The low limit value is set
higher than the high limit
value.
Reset
the
limit
values.
E4
The address of storing unit
weight is out of the address
range.
Input new address.
--OL--
Low battery charge.
Recharge the battery.
Load exceeds scale capacity.
Reduce the load on
the pan.
A value exceeds the display
range.
Cancel
some
operation item.
--------
System is busy.
Just wait.
If the error message still is shown after above ways, please
recalibrate. If the problem still cannot be solved then contact your
dealer for further support.