HIGH PRECISION ELECTRONIC SCALE
Operation Manual
JS-AZ
Page 1: ...HIGH PRECISION ELECTRONIC SCALE Operation Manual JS AZ...
Page 2: ...ity accumulation 15 viii Preset counting check range 16 ix Preset weight check range 16 VI User Programming Functions 20 i Auto shut off time span 20 ii Change unit of weight 20 iii Backlight type 21...
Page 3: ...ity extremes These factors will affect displayed weight reading DO NOT install the scale Next to open windows or doors causing drafts or rapid temperature changes Near air conditioning or heating vent...
Page 4: ...h a level indicator on the left bottom side of the front panel and four adjustable leveling feet Adjust the leveling feet until the bubble appears in the center circle of the indicator Turn on Scale D...
Page 5: ...D Display Windows Weight Display Total 6 digits for weight accumulated or being measured on the pan Unit Weight Display Total 6 digits for unit weight or times of weight accumulated Total Pieces Displ...
Page 6: ...ator disappears U W lack Lack of Unit Weight If the unit weight is less than 1 10 display divisions a triangular indicator will appear to remind the user that the displayed unit weight is too small fo...
Page 7: ...ect the desired value in setting mode SMPL Use this key to input sample size U WT SAVE Use this key to input the known unit weight of item to be counted Use this key to store unit weight into numeric...
Page 8: ...RS 232 which can be connected to PC printer and a control box to output 3 section control signal 2 9 pins connector for a remote platform Position of the connectors Both of the connectors are located...
Page 9: ...emote platform connector RS232 Connector 3 color alarm lamp 1 6 2 7 3 8 4 9 5 PIN 1 Excitation 3 Shield 5 Excitation 7 Signal 9 Signal 1 6 2 7 3 8 4 9 5 PIN 2 RXD 3 TXD 5 GND 6 VCC 7 Yellow 8 Green 9...
Page 10: ...er the remote platform can be used WEIGHT UNIT WEIGHT COUNT 0 0000 0 0 Ib kg g Local Remote HI OK LO ADD Press LOCAL REMOTE key to return to local platform WEIGHT UNIT WEIGHT COUNT Change to local Ib...
Page 11: ...eturn to normal counting mode WEIGHT UNIT WEIGHT COUNT 0 0 0 0 Ib kg g Local Remote HI OK LO ADD Note 1 The default platform is Local platform 2 The default weighting unit of local platform is g while...
Page 12: ...ocal Remote HI OK LO ADD Sample Weight 2 Input the quantity of item on the pan Value in COUNT window will blink WEIGHT UNIT WEIGHT COUNT 2000 0 50 40 Ib kg g Local Remote HI OK LO ADD Sample Weight In...
Page 13: ...ght during counting process Known unit weight 1 Input the known unit weight Value in COUNT window will blink WEIGHT UNIT WEIGHT COUNT 0 0 40 0 Ib kg g Local Remote HI OK LO ADD Known unit weight 2 Pre...
Page 14: ...IGHT COUNT 0 0 Store 0 Ib kg g Local Remote HI OK LO ADD 3 Enter an address cell 000 200 total 200 cells available by pressing any of the numeric keys 0 9 then press the U WT SAVE key to store the uni...
Page 15: ...te HI OK LO ADD 3 The scale will enter into counting mode while stable display appears as below WEIGHT UNIT WEIGHT COUNT 0 0 0 0 Ib kg g Local Remote HI OK LO ADD Container s weight known Nothing on t...
Page 16: ...GHT COUNT 960 0 0 Ib kg g Local Remote HI OK LO ADD Eliminate TARE Remove all on the pan the weight display will show a negative container s weight Pressing TARE P TARE MOVE key at this moment will br...
Page 17: ...5 Ib kg g Local Remote HI OK LO ADD Total accumulated Total accumulation Total accumulated Weight times count Accumulation effective only when stays at zero 4 Press TOTAL PRINT ENTER key or wait appr...
Page 18: ...mulation status mode and press CLEAR EXIT key to clear all accumulated data VIII Preset counting check range Users can set a Hi Lo range for counting check when the number of objects on the pan is wit...
Page 19: ...NT lo 2000 5 Ib kg g Local Remote HI OK LO ADD Low limit High limit value keyed in Low limit value keyed in 4 Press SMPL key to complete counting check range preset procedures and return to normal cou...
Page 20: ...D High limit High limit value keyed in Window for displaying low limit value to be input 3 Press ALARM key again and key in the desired low limit value as indicated below low limit value effective onl...
Page 21: ...lease refer to Page20 II of section V Backlight typesetting When the backlight type is set to be Auto there are three colors for check weight count Red color The weight count on the pan is higher than...
Page 22: ...Press TARE P TARE MOVE key to revolve the system preset time span 2min 5 min 8 min and 0 Default setting 0 3 Press CLEAR EXIT key to determine and return to normal counting mode or press TOTAL PRINT E...
Page 23: ...g Local Remote HI OK LO ADD Backlight type Backlight Internal count 2 Press TARE P TARE MOVE key to revolve the system preset backlight type 0 auto backlight 1 manual backlight Default setting 0 3 Pre...
Page 24: ...GHT COUNT 0 RU 82483 Ib kg g Local Remote HI OK LO ADD Recomputing mode Averaging Internal count 2 Press TARE P TARE MOVE key to revolve the system preset Recomputig modes Default setting 0 0 disable...
Page 25: ...0 trn 82483 Ib kg g Local Remote HI OK LO ADD Transmit mode type Transmit mode Internal count 2 Press TARE P TARE MOVE key to revolve the system preset transmit mode types Default setting 0 0 output d...
Page 26: ...EAR EXIT key to determine and return to normal counting mode or press TOTAL PRINT ENTER key for determination and move to next VII Label format available when a label printer is connected Only when tr...
Page 27: ...R key in USER PROGRAMMING FUNCTION MODE and release until the following displays appear WEIGHT UNIT WEIGHT COUNT 0 PASS 82483 Ib kg g Local Remote HI OK LO ADD check alarm type check alarm internal co...
Page 28: ...ight or total count falls outside the set range Ex 1 Counting check alarms Quantity in COUNT window blinks WEIGHT UNIT WEIGHT COUNT 2000 0 20 100 Ib kg g Local Remote HI OK LO ADD Over high limit Ex 2...
Page 29: ...d 4 Default setting 2 The larger number selected the wider range 3 Press CLEAR EXIT key to determine and return to normal counting mode or press TOTAL PRINT ENTER key for determination and move to nex...
Page 30: ...R PROGRAMMING UNCTION MODE and release until the following displays appear WEIGHT UNIT WEIGHT COUNT 1 filt 82483 Ib kg g Local Remote HI OK LO ADD Stable class range Stable class Internal count 0 5 2...
Page 31: ...HI OK LO ADD Zero return range Zero return Internal count 0 5 2 Press TARE P TARE MOVE key to revolve the system preset zero return range 0 1 2 3 4 5 Default setting 0 The larger number selected the...
Page 32: ...Press ADD key to confirm The displayed reading in the UNIT WEIGHT window starts blinking The scale will stop blinking and return to normal counting mode The calibration is now completed Required weig...
Page 33: ...wer supply module to the scale and plug in The scale does not need to be turned on The battery should be charged for 12 hours for full capacity There is an LED to indicate the status of battery chargi...
Page 34: ...p bit 1 BIT Code format Code ASCll Connector 9 Pin Socket Pin2 Input Pin3 Output Pin5 Signal Ground 3 Transmit Format when it is in Accumulation mode and transmit by pressing the following keys Press...
Page 35: ...7 byte TNB Total net weight no dot 6 byte TTA Total Tare weight 7 byte TTB Total Tare weight no dot 6 byte TGA Total Gross weight 7 byte TGB Total Gross weight no dot 6 byte UWA Unit weight 7 byte UWB...
Page 36: ...remote platform properly to the scale E3 The low limit value is set higher than the high limit value Reset the limit values E4 The address of storing unit weight is out of the address range Input new...
Page 37: ...low and green LED backlight the colors are changeable in check weight count mode to indicate different check status Zero Range 4 Tare Range Full Capacity by Subtraction Stabilization Time 2 seconds Op...