background image

©JRI 

 

VI.

 

BATTERY REPLACEMENT 

Removing the battery 

Open the battery cover 

 

with a suitable 

object (coin) to align the marks  

(/!\0 = Open ; 1= Closed) 

Remove the battery 

 from its lodging 

 

Replacing the battery 

Put the new battery 

 in place respecting 

the polarity 

.  

Battery detection is confirmed by the 

activation of the red LED 

 for few seconds. The device can be activated after the extinction of the red LED. 

 

 

 

*Recommended batteries: Saft LS14500 type AA 3.6V 2250mAh 

 

VII.

 

MAINTENANCE 

Clean the device with a soft cloth, either dry or slightly moistened with water. To remove stubborn dust, 

use a cloth soaked in a diluted, non-abrasive detergent. Then wipe carefully with a soft dry cloth.  

Never use benzene, thinner, alcohol or any type of solvent, which can cause discoloration or deformation 

of the surfaces. 

VIII.

 

TECHNICAL FEATURES 

a)

 

Common features: 

HMI 

: 1 RGB LED + 1 sensitive bouton 

Frequency band 

: 2.4GHz (from 2400 to 2483.5 MHz) 

Maximum Radio Power  

: 10 dBm 

Memory 

: 10 000 timestamped measures  

Resolution 

: 0.01 

Dimensions 

: 63 mm x 42 mm x 25 mm 

Case 

: Polycarbonate – Food Contact  

Power 

: 3,6v Lithium battery up to 6 years life time battery, 

depending on usage 

Weight 

: ~ 60g 

 

b)

 

Specific features: 

Nano SPY T1 Temperature (internal probe)

 

 

 

 

Sensor 

: PT100 sensitive element inside unit 

Operating range 

: -40+85°C 

Measurement range 

: -40+85°C 

Accuracy, standard version 

: ±0.4°C from -20°C to +40°C / ± 0.5°C outside this range  

IP rating 

: IP 68 

Frequency of measurement and 

transmission 

: adjustable from 1 min to 12h 

Frequency of recording 

: adjustable from 1 min to 24h 

Response time 

: ~ 5 min. to 90% of the variation 

 

 

KEEP THE BATTERY AWAY FROM FIRE; DO NOT ATTEMPT TO RECHARGE OR SHORT-CIRCUIT IT 

ONLY USE BATTERIES* SUPPLIED BY JRI (REF : 11596) 

 

 

 

 

 

   

 

 

 

 

Summary of Contents for Nano SPY

Page 1: ...USER GUIDE Nano SPY Ref 11816J...

Page 2: ...CRIPTION 4 A CONTROL UNIT 4 B MOUNTING 4 IV OPERATION 5 A OFF STATE 5 B ACTIVATION 5 C TURNING OFF 5 D ACTIONS ON THE TOUCH BUTTON 5 V NANO SPY U CONNECTION 6 VI BATTERY REPLACEMENT 7 VII MAINTENANCE...

Page 3: ...not dispose of in a refuse dump or waste disposal bin Comply with existing legislation for disposal Power source this device is powered by a 3 6VDC type AA lithium battery ch V CE LABELING this device...

Page 4: ...Nano SPY LINK on the Ethernet network To ensure your safety during installation or an intervention on a device placed in a high position use proper equipment which is in good condition and provides ad...

Page 5: ...Sirius if it is contact with a Link It starts to measure and transmit its measurements to My Sirius at the frequency defined in MySirius then flashes regularly as a function of its status c Turning Of...

Page 6: ...be disconnected from the logger for replacement or for the exchange of the logger itself An IP67 Binder connector optional can be used to facilitate the connection of the sensors Connect only sensors...

Page 7: ...se discoloration or deformation of the surfaces VIII TECHNICAL FEATURES a Common features HMI 1 RGB LED 1 sensitive bouton Frequency band 2 4GHz from 2400 to 2483 5 MHz Maximum Radio Power 10 dBm Memo...

Page 8: ...ng range from 30 70 C Measurement range 50 105 C Accuracy standard version 0 3 C from 20 to 30 C 0 5 C outside this range Accuracy incubator version 0 2 C from 30 to 50 C 0 5 C outside this range IP r...

Page 9: ...20 C to 50 C 0 3 C from 80 C to 20 C and from 50 C to 140 C 0 5 C beyond these ranges Current input measurement range resolution accuracy control unit only No detection of probe failure for 0 20mA in...

Page 10: ...ance statement can be requested by email info group mms com IC CANADA This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 Thi...

Page 11: ...JRI 11 IX OPERATION CAPACITY...

Page 12: ...JRI 12...

Page 13: ...by offering you a fully integrated solution This global service offer includes both maintenance and a metrological service ensuring the optimum functioning of your devices or of your installation You...

Reviews: