background image

©JRI 

 

I.

 

INTRODUCTION 

The Nano SPY is a measuring device for 1 or 2 physical parameters (T or TH depending on the model). Data is 

transmitted through NANO Link or Relay/Alarm modules using wireless 2.4GHz radio signal to the JRI-MySirius 

monitoring software hosted on the JRI could or the customer’s server.  

  

The Nano SPY complies with EN 12830 with temperature probes only, and is compatible  

with EN 13486 which defines procedures for periodic verification. 

 

a)

 

Product contents 

 

1 Nano SPY  

 

1 User guide 

 

b)

 

Symbols  

 

Do not use the device under conditions other than those described in the technical 

characteristics 

Risk of fire or explosion in the case of improper use: 

- Recharging of the battery 

- Short circuiting of the battery 

If the device is used in a manner not specified by the manufacturer, the protection 

provided by the device may be compromised. 

 

II.

 

INSTALLATION RECOMMANDATIONS 

To ensure optimal radio transmission, a certain number of recommendations must be respected, as any 

wireless transmission is subject to disturbances. 

a)

 

Sources of disturbances or attenuation 

 

The presence of obstacles in the wave path between the Nano SPY and the Nano SPY Link (wall, 

furniture, people…) or near the antenna. 

 

The thickness of an obstacle in the wave path. The attenuation is greater diagonally than 

perpendicularly. 

 

 

 

 

RECYCLING: do not dispose of in a refuse dump or waste disposal bin. Comply with existing legislation for disposal. 

 

Power source:  this device is powered by a 3.6VDC type AA lithium battery (§ ch. V). 

 

CE LABELING: this device is certified to conform to European regulations for electrical safety, flammability, disruptive 

electromagnetic emissions, and immunity to environmental electrical disturbances. 

 

FCC ID : W45 12525 

This device complies with part 15 of the FCC rules. Operation is subject to the following two condition:  

1.

 

This device may not harmful interference, 

2.

 

This device must accept any interference received, including interference that may cause undesired operation. 

The grantee is not responsible for any changes or modification not expressly approved by the party responsible for compliance. 

Such modifications could void the user’s authority to operate the equipment. 

NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the 

FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is 

operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed 

and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this 

equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the 

interference at his own expense.

 

Summary of Contents for Nano SPY

Page 1: ...USER GUIDE Nano SPY Ref 11816J...

Page 2: ...CRIPTION 4 A CONTROL UNIT 4 B MOUNTING 4 IV OPERATION 5 A OFF STATE 5 B ACTIVATION 5 C TURNING OFF 5 D ACTIONS ON THE TOUCH BUTTON 5 V NANO SPY U CONNECTION 6 VI BATTERY REPLACEMENT 7 VII MAINTENANCE...

Page 3: ...not dispose of in a refuse dump or waste disposal bin Comply with existing legislation for disposal Power source this device is powered by a 3 6VDC type AA lithium battery ch V CE LABELING this device...

Page 4: ...Nano SPY LINK on the Ethernet network To ensure your safety during installation or an intervention on a device placed in a high position use proper equipment which is in good condition and provides ad...

Page 5: ...Sirius if it is contact with a Link It starts to measure and transmit its measurements to My Sirius at the frequency defined in MySirius then flashes regularly as a function of its status c Turning Of...

Page 6: ...be disconnected from the logger for replacement or for the exchange of the logger itself An IP67 Binder connector optional can be used to facilitate the connection of the sensors Connect only sensors...

Page 7: ...se discoloration or deformation of the surfaces VIII TECHNICAL FEATURES a Common features HMI 1 RGB LED 1 sensitive bouton Frequency band 2 4GHz from 2400 to 2483 5 MHz Maximum Radio Power 10 dBm Memo...

Page 8: ...ng range from 30 70 C Measurement range 50 105 C Accuracy standard version 0 3 C from 20 to 30 C 0 5 C outside this range Accuracy incubator version 0 2 C from 30 to 50 C 0 5 C outside this range IP r...

Page 9: ...20 C to 50 C 0 3 C from 80 C to 20 C and from 50 C to 140 C 0 5 C beyond these ranges Current input measurement range resolution accuracy control unit only No detection of probe failure for 0 20mA in...

Page 10: ...ance statement can be requested by email info group mms com IC CANADA This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 Thi...

Page 11: ...JRI 11 IX OPERATION CAPACITY...

Page 12: ...JRI 12...

Page 13: ...by offering you a fully integrated solution This global service offer includes both maintenance and a metrological service ensuring the optimum functioning of your devices or of your installation You...