7. Operation
The platelet stirrer has been designed for a quiet and continuous blood bag operation.
To enable a continuous operation monitoring, the equipment records all break times,
both due to loading and unloading, as unforeseen ones, due to power failure.
7.1 Starting-up
Start the equipment by pressing the main switch.
By a (short) press of the button, it stops for loading and unloading.
The stop lasts for a determined «T» time. After this time, the stirrer starts automatically.
With a long press you can access the record menu. See chapter 8.
7.2 Break counter
One can access the break counter by the single button (chapter 8).
This counter can be reset to zero.
The break lasts for a determined «T» time. After this time, the stirrer starts auto-
matically.
7.3 Power supply fails
The computer registers the power failures.
For every stop, the computer registers date, time and duration of the stop.
Registration of stops can be reset.
7.4 Password
Access to the above parameters is password protected.
Password is 1111.
Password can be momentarily unlocked, to facilitate access to counters and registers
before initiating treatment.
7.5 Settings
The agitation speed is factory set. If it is necessary to modify this speed, remove
the blue rear plug and adjust the speed by using a flat screwdriver.