ULTRATERMOSTATO DE INMERSIÓN

IMMERSION ULTRATHERMOSTAT

 

    ULTRATERM TFT -200 S  

3000554           

 

    ULTRATERM TFT -200 

3000545

Summary of Contents for 3000545

Page 1: ...CTA s a u Autov a A 2 Km 585 1 Abrera 08630 Barcelona Espa a Tel 34 937 700 877 Fax 34 937 702 362 e mail selecta jpselecta es website http www jpselecta es ULTRATERMOSTATO DE INMERSI N IMMERSION ULTR...

Page 2: ...a temperatura 7 7 2 Pantalla de AJUSTES 8 7 3 Pantalla de programaci n 8 7 4 Seleccionar un programa para ejecutarlo desde la pantalla de Trabajo 9 7 5 Pantalla de Trabajo 9 7 6 Bomba de circulaci n 9...

Page 3: ...display 0 1 C 0 1 C Funcionamiento temporizado Horas minutos Horas minutos Caudal bomba 10L min 10L min Mueble en acero pintado epoxy en acero inox con asas de transporte laterales 1 Informaci n gener...

Page 4: ...peratura 4 Informaci n sobre seguridad en la utilizaci n del equipo 4 4 Mantenimiento Antes de manipular en su interior desconecte el equipo de la toma de red La manipulaci n de los circuitos electr n...

Page 5: ...el ctrica Escoger una toma de corriente cercana al equipo y adecuada a la potencia del equipo Utilizar el cable de conexi n suministrado u otro de similares ca racter sticas Para su seguridad la toma...

Page 6: ...LA TRABAJO PROGRAMA 00 SAVE TIEMPO 00 00 TEMP 16 0 C OVERTMP 0 C UNDERTMP 0 C SETUP HH MM SS DD MM YY HH MM SS DD MM YY 22 9 C SET 35 2 FLOW 80 DURACI N SET H M 0 20 40 80 60 20 40 Anticong coche Car...

Page 7: ...H MM SS DD MM YY 22 9 C SET 40 0 FLOW 80 DURACI N SET H M 3Seleccionar la temperatura con las teclas su bir y bajar En el ejempo 40 C HH MM SS DD MM YY 22 9 C SET 40 0 FLOW 80 DURACI N SET H M HH MM S...

Page 8: ...ecuencia de toma de datos para el registro b Sensor externo Seleccione el valor Si si desea trabajar con una sonda externa conectada al conector de la parte posterior del equipo c Calibraci n OFFSET S...

Page 9: ...uraciones Circulaci n INTERNA El agua sale de la bomba por el tubo horizontal Ver fig 7 6 1 Con esta configuraci n se consigue una buena homogeneizaci n de la temperatura en el l quido de la cubeta Ci...

Page 10: ...el bajo de temperatura A 04 External RTD opened Sonda externa abierta A 05 External RTD shortcir cuited Sonda externa cruzada A 06 Internal RTD opened Sonda interna abierta A 07 Internal RTD shortcir...

Page 11: ...ONES CODIGO 80333 REV B 07 2020 Pag 11 J P SELECTA s a u Autov a A 2 Km 585 1 Abrera 08630 Barcelona Espa a Tel 34 937 700 877 Fax 34 937 702 362 e mail selecta jpselecta es website http www jpselecta...

Page 12: ...e any doubts or enquiries contact with J P SELECTA s technical service Do not use the equipment in explosive flammable or corrosive atmospheres 3 Technical Specification 3000545 3000554 Dimensions Hei...

Page 13: ...eat flammable or explosive materials or those which chemically react with violence Hazard risk Electric risk Risk of high temperature contact 4 Information about safety in the use of the equipment 4 4...

Page 14: ...he drain valve is closed 6 2 Connection to the mains Plug the equipment to a near socket suitable for the equipment power Use the connection cable supplied or a similar one For your safety the equipme...

Page 15: ...OPERATION SCREEN PROGRAM 00 SAVE TIME 00 00 TEMP 16 0 C OVERTMP 0 C UNDERTMP 0 C SETUP HH MM SS DD MM YY HH MM SS DD MM YY 22 9 C SET 35 2 FLOW 80 DURATION SET H M 0 20 40 80 60 20 40 Anticong coche C...

Page 16: ...s over HH MM SS DD MM YY 22 9 C SET 40 0 FLOW 80 DURATION SET H M 3Set the temperature with and 40 C on the picture HH MM SS DD MM YY 22 9 C SET 40 0 FLOW 80 DURATION SET H M HH MM SS DD MM YY 22 9 C...

Page 17: ...sensor Select the value YES if you want to work with an external probe connected to the connector on the back side of the equipment c Calibration OFFSET Use only this setting if you notice a deviation...

Page 18: ...circulation pump could operate in two configuration modes INTERNAL circulation Water or other will be pumped from horizontal pump output See fig 7 6 1 Use this configuration to make water circulation...

Page 19: ...rature level A 04 External RTD opened Open external probe A 05 External RTD shortcircuited Shortcircuited external probe A 06 Internal RTD opened Open internal probe A 07 Internal RTD shortcircuited S...

Page 20: ...ODE 80333 REV B 07 2020 Page 20 J P SELECTA s a u Autov a A 2 Km 585 1 Abrera 08630 Barcelona Espa a Tel 34 937 700 877 Fax 34 937 702 362 e mail selecta jpselecta es website http www jpselecta es J P...