- 5 -
Chapter 1 Overview
1.1 Scope of application
It is applicable to accurate, constant and continuous intravenous infusion in medical institutions.
1.2 Technical Parameters
IV Set Options /Size
: Open System for common widely-used brand of 10d/ml
、
15d/ml
、
20d/ml
、
60d/ml
(3.4mm
~
4.5mm), Also can be customized for dedicated IV set
Infusion Mode:
KVO:
Adjustable from 1ml/h - 20ml/h
Infusion Rate
: 20 d/ml IV set
:
0.1
~
1200ml/h
;
1.0
~
400d/min
60 d/ml IV set
:
0.1
~
150ml/h
;
1.0
~
150d/min
0.1 ml/h increments; 1d/min increments
Infusion mode
Description
Operation
Remark
Volume mode
(
ml/h
)
Choose mode needed
VTBI/time/rate
VTBI / rate
VTBI / time
Time / rate
Simple rate mode
Rate:0.1 ~1200ml/h
VTBI
:
0~9999ml
Time
:
1min~99h59min
VOL
:
0.1~9999ml
Drop mode
(
Drops/min
)
Only For Option
Choose mode needed
VTBI / time / rate
VTBI / rate
VTBI / time
Time / rate
Simple rate mode
Rate: 1d/min ~ 400d/min
(
10d/ml
、
15d/ml
、
20d/ml
)
Rate: 1d/min ~ 150d/min
(60 d /ml)
VTBI
:
0~9999ml
Time
:
1min~99h59min
VOL
:
0.1~9999ml
Program mode
Choose mode needed
VTBI / rate
Pause time
8 stages of programming
Can stop infusion to program
Weight mode
Choose mode needed
Dose/weight
Drug Library mode
Choose mode needed
Dose/ rate
Rate
Can import drug library
49 kinds of drugs inside
machine
Bolus
Automatic
VTBI / rate
Rate
:
0.1~1200ml/h adjustable
VTBI
:
1~9999ml adjustable
Default 600ml/h
,
2ml
Manual
VTBI / rate
KVO
Opens automatically as
infusion is completed
Whether displays KVO
volume and rate setting
1~20ml/h adjustable
1ml/h default