TROUBLESHOOTING
SPECIFICATIONS
PROBLEM
CAUSE ANALYSIS
SOLUTION
FLOWMETER WILL
NOT TURN OFF
Inspect needle valve for
wear. Check adjustments.
Replace needle valve
or make adjustments.
ANIMALS WILL NOT GO TO
SLEEP/STAY ASLEEP
1. Inspect system for leak
2. Inspect breathing circuit
and bag for leak or loose
connections.
3. Check active waste
gas system function/
attenuation.
4. Confirm vaporizer has
an adequate amount of
agent.
1. Tighten any loose
connections (vaporizer
endcaps, tubing
connections).
2. Replace breathing
circuit or bag. Fix loose
connections.
3. Ensure active waste
gas system function/
attenuation is working
properly
4. Fill vaporizer with correct
amount of agent.
FOR ASSISTANCE, CALL 1.800.525.5614
Weight 45 lbs./20.5 kg
Dimensions 21”W x 21”D x 53”H
(53mm x 53mm x 135mm)
Vaporizer Endcaps 23mm tapered
Flowmeter 0.2-4 LPM
Oxygen (Carrier Gas) Inlet DISS Male
Frame Aluminum