Johnson Health Tech. Co., Ltd.
The probe is
placed on the
power supply
input end
Switch to AC,
and turn on the
power supply to
confirm voltage
Test if the output
end voltage
conforms to the
input end
After the power is
off,
measure if the
fuse is damaged
The main purpose is to measure if the input end voltage has output
voltage (AC110V or AC220V) after passing through the filter; if there
is no voltage, check if the fuse is burned (
The power must be off
when measuring the fuse
)
When the fuse is not burned, and there is still no output voltage,
replace the filter
4-1.
Fail to start
Fail to start
Fuse
specification