background image

2    T-4000 Technical Bulletin

increase or up to decrease
the sensitivity.

5.

Check the output pressure.
If it is different from the value
found in step 3, turn the
sensitivity slider adjusting
screw until the output
pressure equals that value.
For low volume output
(plastic body) instruments,
use the output adjusting
screw to adjust the output
pressure.

6.

Check the sensitivity.
Change the control setting
on the dial by at least 5F

°

and note the new output
pressure.  The change in
output pressure divided by
the change in dial setting is
the actual sensitivity.

7.

Repeat steps 4 through 6
until the desired sensitivity is
obtained.

T-4003 Range of Remote
Readjustment

The range of remote
readjustment is adjustable from
.15 to 2.5 F

°

/PSI (.012 to

.2C

°

/kPa) and is factory set at

0.5 F

°

/PSI (.04 C

°

/kPa).  To

determine if the range of remote
readjustment needs to be
adjusted, proceed as follows:

1.

Insert the test gage into the
test port fitting and note the
output pressure.

2.

Set the master pressure to
4 psig (28 kPa).

3.

Turn the set point dial to 80F
and turn the calibration
screw to obtain a 3 psig
(21 kPa) output pressure.

4.

Set the master pressure to
14 psig (98 kPa).

5.

Turn the set point dial until
the output pressure returns
to 3 psig.

6.

The change in dial setting
divided by the 10 psig
(70 kPa) change in master
pressure (14 psig - 4 psig =
10 psig) is the actual range
of readjustment.  If this is not
the desired range, move the
range of readjustment slider
(sensitivity slider in Fig. 1a)
toward the control port to
decrease or away from the
control port to increase the
range of readjustment.
Readjust the slider screw to
the pressure noted in step 1
above.  Repeat steps 2
through 6 until the desired
range is obtained.

Set Point Adjustment

Before calibrating the instrument,
check to see if the output
pressure is already at an
appropriate value according to
the ambient temperature.  For

example, if the actual room
temperature is below the set
point and the output signal is
calling for more heat, the
thermostat is doing its job and
does not require adjustment.

T-4002 and T-4100 (See Fig. 1)

1.

Insert the test gage.

2.

Furnish supply air to the
instrument (restricted supply
for T-4100).

3.

Note the ambient
temperature in the space.

4.

Turn the set point dial to that
temperature.

5.

Turn the calibration screw
until the output pressure is at
the mid spring range of the
controlled device.

6.

Turn the set point dial to the
desired set point.

T-4003 Set Point Adjustment

Insert the gage and set the
master pressure to one half of
the supply pressure.  Follow
steps 3 through 6 under T-4002
Set Point Adjustment.

Archived Document

Summary of Contents for T-4002

Page 1: ...in test port fitting which accepts a hypodermic needle test probe and gage X 200 140 and G 2010 11 On older models the test port cap screw must be removed to accept a screw in test gage X 200 19 For these units a test port fitting T 4000 615 can be ordered All low volume output thermostats have hypodermic needle test points For units that require a screw in gage remember to replace the test port c...

Page 2: ...sure returns to 3 psig 6 The change in dial setting divided by the 10 psig 70 kPa change in master pressure 14 psig 4 psig 10 psig is the actual range of readjustment If this is not the desired range move the range of readjustment slider sensitivity slider in Fig 1a toward the control port to decrease or away from the control port to increase the range of readjustment Readjust the slider screw to ...

Page 3: ...roceed as follows 1 Remove the T 4054 from the mounting bracket and pull it away from the wall so the deadband adjusting screw on the back of the instrument is exposed see Fig 3 2 Turn the screw clockwise to increase or counter clockwise to decrease the deadband pressure until the gage reads the desired pressure Approximately one half turn will produce a 1 PSI 7 kPa change in deadband pressure 3 R...

Page 4: ...use the proper element and supply pressure for the particular application The T 4506 is furnished with DAY and NIGHT labels which can be applied to the appropriate element Day Set Point Adjustment See Fig 5 1 Insert the test gage 2 Furnish 15 psig supply pressure for day control lower dial 3 Note the ambient temperature in the space 4 Turn the lower set point dial to that temperature 5 Turn the lo...

Page 5: ...ture in the space 4 Turn the lower set point dial to that temperature 5 Turn the lower calibration screw on the heating element until the output pressure is at the mid spring range of the controlled device 6 Turn the lower set point dial to the desired heating set point 5 Turn the calibration screw on that element until the output pressure to the heating apparatus is at the mid spring range of the...

Page 6: ...e supply pressure to 15 psig and rotate the lower dial to produce approximately 5 psig 35 kPa on the output gage 4 Increase the supply pressure to 20 psig and rotate the upper dial to produce approximately 15 psig on the output gage 5 Set the supply pressure to the desired switchover point 6 Slowly turn the switchover adjusting screw see Fig 8 clockwise to increase or counterclockwise to decrease ...

Page 7: ...T 4000 Technical Bulletin 7 Notes Archived D ocum ent ...

Page 8: ...8 T 4000 Technical Bulletin Notes Controls Group 507 E Michigan Street P O Box 423 Milwaukee WI 53202 Printed in U S A Archived D ocum ent ...

Reviews: