Using in open procedures
1.
Press the
OPEN
button to switch to
OPEN
mode.
3.
Adjust
RUN TIME
settings to
user-preferred time-delay intervals.
RUN TIME
options control how
long flow continues after the
energy device is deactivated when
utilizing one of the automatic
activation methods. This feature
enables evacuation of residual
smoke after the surgeon discon-
tinues activating the energy device.
2.
Adjust
FLOW
settings to user-
preferred flow-rate levels.
4.
Ensure smoke evacuation tube
is connected to MEGADYNE™
Smoke Evacuator.
118LPM
106LPM
94LPM
82LPM
71LPM
30 sec.
15 sec.
10 sec.
6 sec.
3 sec.
4.
Insert the filter securely into the
filter port located on the front of the
MEGADYNE™ Smoke Evacuator.
6.
Check the
FILTER
life indicator:
5.
Turn on the machine: Press the
power switch button to turn on the
smoke evacuator.
If the filter is not securely seated,
all filter lights will blink until proper
installation is completed.
One indicator light will be
illuminated when the filter is secure
and indicate remaining filter life.
a.
Green
Ready for use:
100% - 25% filter life.
b.
Yellow
Ready for use:
25% - 0% filter life.
c.
Red
Once the red light is illuminated,
the smoke evacuator can only be
activated within a 3 hour window
to complete the case.
Before you begin (cont.)