Using in open procedures

1.

   Press  the 

OPEN

 button to switch to

OPEN

 mode.

3.

 

Adjust 

RUN TIME

 settings to

user-preferred time-delay intervals.

 
 
 
 
 

RUN TIME

 options control how 

long flow continues after the 
energy device is deactivated when 
utilizing one of the automatic 
activation methods. This feature 
enables evacuation of residual 
smoke after the surgeon discon-
tinues activating the energy device.  

2.

   Adjust 

FLOW

 settings to user-

preferred flow-rate levels.

4.

 Ensure smoke evacuation tube
is connected to MEGADYNE™
Smoke Evacuator.

118LPM 

106LPM 

94LPM 

82LPM 

71LPM

30 sec. 

15 sec. 

10 sec. 

6 sec. 

3 sec.

4.

 Insert the filter securely into the
filter port located on the front of the
MEGADYNE™ Smoke Evacuator.

6.

Check the 

FILTER

 life indicator: 

5.

 Turn on the machine: Press the
power switch button to turn on the
smoke evacuator.

If the filter is not securely seated,  
all filter lights will blink until proper 
installation is completed.

One indicator light will be 
illuminated when the filter is secure 
and indicate remaining filter life.

a.  

Green 

Ready for use:
100% - 25% filter life.

b.  

Yellow 

Ready for use:
25% - 0% filter life.

c.  

Red 

Once the red light is illuminated,
the smoke evacuator can only be
activated within a 3 hour window
to complete the case.

Before you begin (cont.)