AT TA C H T H E P U M P C AT H E T E R
Aseptically deliver the pump and pump catheter to a
sterile surface.
Firmly press the catheter’s silicone connector onto the
pump’s catheter port.
Secure the connector with a heavy, non-absorbable
suture placed in the suture groove at the base of
the connector.
Check the catheter for any tears.
A LT E R N AT I V E T E C H N I Q U E :
The pump catheter may be attached to the intraspinal
catheter before attaching to the pump.
F I L L A N D P R I M E T H E P U M P
Using the OR Prep Kit, follow the MEDSTREAM Pump
Instructions for Use to perform the following steps.
1. Empty the reservoir
2. Flush the reservoir
3. Fill the reservoir
4. Prime the bolus channel
Note:
Also refer to the MEDSTREAM OR Poster for
detailed instructions and illustrations.
STEP 2.
Prepare the Pump
10
492192.QX7:Layout 1 4/14/09 4:09 PM Page 10