4.  Sequentially reduce the 

Comfort Adjust

 setting 

minimal resistance between the patient’s 

sacral area/buttocks and the lower safety 

mattress. This is the proper 

Comfort Adjust

 

of the air cells and prevent bottoming out of 

the mattress.

5.  Document the patient’s 

Comfort Adjust

  

setting for future reference, and re-evaluate 

with the Hand Check as the patient’s  

condition warrants.

Remove the existing mattress from the bed.
Place the Arise LAL mattress with the hose 
connection at the foot end of the bed and the 
therapy cells facing up. Secure the six (6) straps 
on the mattress securely to the movable part of the 
bed frame.
If the therapy pad is not already on the mattress, 
attach it securely to the mattress. 
Hang the control unit on the foot of the bed facing 
away from the bed. Attach the hose connector 
marked CPR to the underside of the control unit.
Plug the Fowler Boost sensor into the side of the 
control unit. 
Plug in the control unit and the yellow 

Standby 

light will illuminate. Press the 

Power 

key.  

The control unit will start and the green light  
will illuminate.

Select the appropriate 

Comfort Adjust

 setting 

(Refer to “Comfort Adjust Setup” section).

The CairRails risk management side air bolsters 

the turn valve on the hose assembly between the 

the CairRails by moving the turn valve to the up 

Note:

recommended that the support surface be in 

 mode.

Place the Arise LAL overlay on top of the mattress 
with the hose connection at the foot end of the bed 
and the therapy cells facing up. Secure the straps 
around the head and foot ends of the mattress. 
Fasten the two straps by the seat section securely 
to the sleep deck of the bed frame.

If the therapy pad is not already on the overlay, 
attach it securely using the zippers. 
Hang the control unit on the foot of the bed 
facing away from the bed. Attach the hose 
connector marked 

CPR

 to the underside of the 

control unit.
Plug the Fowler Boost sensor into the side of the 
control unit. 
Plug in the control unit and the yellow 

Standby 

light will illuminate. Press the 

Power

 key.  

The control unit will start and the green light  
will illuminate.

Place the patient on the overlay. Overlay can 

depending on patient weight/size.  

Note: 

Keep the control unit on while the patient 

is on the overlay.
The CairRails air side bolsters are currently not 
available with the Arise LAL LAL overlay.

Remove the existing mattress from the bed.
Place the Arise LAL 1000 mattress with the 
hose connection at the foot end of the bed and 
the therapy cells facing up. Secure the ten (10) 
straps on the mattress securely to the movable 
part of the bed frame.
If the therapy pad is not already on the mattress, 
attach it securely to the mattress. 
Hang the control unit on the foot of the bed 
facing away from the bed. Attach the hose 
connector marked 

CPR 

to the underside of the 

control unit.
Plug the Fowler boost sensor into the side of the 
control unit. 
Plug in the control unit and the yellow 

Standby 

light will illuminate. Press the 

Power

 key.  

The control unit will start and the green light  
will illuminate.

Select the appropriate 

Comfort Adjust 

setting 

(Refer to “Comfort Adjust Setup” section).

The CairRails risk management side air bolsters 

the turn valve on the hose assembly between 

Note:

recommended that the support surface be in 

 mode.

Electronic Copy: Reference Only

Notice: Document considered current at time of printing. Verify revision before use. 

Summary of Contents for Arise LAL 1000

Page 1: ...To avoid injury read user s manual before using Electronic Copy Reference Only Notice Document considered current at time of printing Verify revision before use...

Page 2: ...properly within the protrude over the side or between the bed rails when the mattress is being used Danger Explosion Hazard Do not use in the in the presence of smoking materials or open support combu...

Page 3: ...uidelines intended to reduce the risk of entrapment including dimensional limits for critical gaps and spaces between bed system components and clinical guidance for assessment and implementation of b...

Page 4: ...Power Failure CairRails Optional Accessories Troubleshooting 13 Unable to Change Therapy Mode or Adjust Comfort Control Nursing Procedures 13 14 Recommended Linen Changing the Therapy Pad Patient Pos...

Page 5: ...ssure relief by combining low air loss with pulsation Low air loss therapy has been demonstrated to reduce the risk of pressure ulcers as well as being a valuable aid in the treatment of pressure ulce...

Page 6: ...for these patients should be considered on an individual basis and discussed with the attending physician The Arise LAL is comprised of two components Therapy control unit Therapy mattress system per...

Page 7: ...ed low pressure heel section provides lower interface pressures Durable base tub is constructed from 100 heavy weight 1680 denier nylon with a 1 5 oz urethane coating and incorporates bed attachment l...

Page 8: ...poor installation of a ground wire that causes the loss of proper ground at the outlet Inadequate grounding at electrical outlets can occur even if there is a ground wire Wires can become loose over t...

Page 9: ...eat section securely to the sleep deck of the bed frame If the therapy pad is not already on the overlay attach it securely using the zippers Hang the control unit on the foot of the bed facing away f...

Page 10: ...setting Refer to Comfort Adjust Setup section Figure 2 on the lower right of the control CairRails by moving the control knob to the up If the patient needs the extended width functionality of the be...

Page 11: ...d Pulsate Static The unit starts in the Static mode which is the standard low air loss therapy Pulsate Pulsate will slightly decrease the pressure in all cushions every 30 seconds then return to the p...

Page 12: ...nd visual alarm Power Interruption Brown out In the event of a brief power interruption up to twenty 20 minutes the Arise LAL control unit will automatically restart and reset to the previous Comfort...

Page 13: ...ly Special linens are not necessary for the Arise LAL While there is no need for a bottom sheet the therapy pad should be covering the therapy cells at all times The patient should never be lying dire...

Page 14: ...should always be in the lowest practical position when the patient is unattended Make sure areas under and around the frame are clear of objects persons and parts of body before adjusting height Warni...

Page 15: ...y pad can be laundered in a washer and dryer with warm water no more than 120 Fahrenheit A non bleach detergent should be used sparingly Wipe dry or allow to air dry Do not use any steam cleaning devi...

Page 16: ...LAL 1000 Mattress 48 122 cm W 80 203 cm H 8 5 22 cm D Arise LAL 1000EX Mattress 36 42 48 91 107 122 cm W 80 88 203 224 cm L 8 5 22cm D USA 120V AC 60 Hz 4A less than 1A running Operating Conditions A...

Page 17: ...Electronic Copy Reference Only Notice Document considered current at time of printing Verify revision before use...

Page 18: ...e instructions is not covered Joerns Arise LAL contains various parts that wear from normal use Joerns Healthcare s obligation under it s warranty is limited to supplying replacement parts servicing o...