Remove the existing mattress from the bed.
Place the Arise LAL 1000 EX mattress with the 
hose connection at the foot end of the bed and 
the therapy cells facing up. Secure the eight 
(8) straps on the mattress securely to the loops 
provided on the Bari10A bed frame.  

Note:

 These straps should not be over tightened 

as they may interfere with the Bari10A bed deck 
elements during width reduction.
If the therapy pad is not already on the mattress, 
attach it via the longitudinal zippers to the 
mattress. 

Note:

 This will require threading the 

hose set through the control window opening 
on the therapy cover located on the mattress 
side at the patients left foot. Also attach the 
snaps around the control panel opening from 
the therapy cover to the mattress control panel 
in four (4) places so that the control panel is not 
obscured (Figure 2).
Hang the control unit on the foot of the bed 
facing away from the bed. Attach the hose 
connector marked 

CPR

 to the underside of the 

control unit.
Plug the Fowler boost sensor into the side of the 
control unit. 
Plug in the control unit and the yellow 

Standby 

light will illuminate. Press the 

Power 

key.  

The control unit will start and the green light  
will illuminate.

Select the appropriate 

Comfort Adjust

 setting 

(Refer to “Comfort Adjust Setup” section).

(Figure 2) on the lower right of the control 

CairRails by moving the control knob to the up 

If the patient needs the extended width 
functionality of the bed (36", 42" or 48") then 
complete the following procedure. 

Note:

 this can 

be accomplished with the patient on the bed:
1.  Adjust the Bari10A bed frame with the knee 

2.  Widen the bed frame (see bed manual) 
3.  Un-strap the mattress side straps from the 

Bari10A bed frame. 

4.  Un-snap the pocket that contains the 

extension side cells so they are free  

5.  Attach the extended width mattress side 

straps to the extended bed deck (

Note:

 do not 

over tighten).

6. 

 

the lower left knob on the mattress control 
panel up. After 2-5 minutes the cells will 

If the patient needs the extended length 
functionality of the bed (80" - 88" then complete 
the following procedure. 

Note:

 this can be 

accomplished with the patient on the bed:

1.  Adjust the Bari10A bed frame with the knee 

2.  Extend the bed frame (see bed manual) 
3.  Un-strap the mattress foot straps from the 

Bari10A bed frame. 

4.  Un-snap the pocket that contains the 

extension foot cells so they are free  

5.  Attach the extended length mattress foot 

straps to the extended bed deck (

Note:

 do 

not over tighten).

6. 

lower left knob on the mattress control panel 
(Figure 2) up. After 2-5 minutes the cells will 

Figure 2

Electronic Copy: Reference Only

Notice: Document considered current at time of printing. Verify revision before use. 

Summary of Contents for Arise LAL 1000

Page 1: ...To avoid injury read user s manual before using Electronic Copy Reference Only Notice Document considered current at time of printing Verify revision before use...

Page 2: ...properly within the protrude over the side or between the bed rails when the mattress is being used Danger Explosion Hazard Do not use in the in the presence of smoking materials or open support combu...

Page 3: ...uidelines intended to reduce the risk of entrapment including dimensional limits for critical gaps and spaces between bed system components and clinical guidance for assessment and implementation of b...

Page 4: ...Power Failure CairRails Optional Accessories Troubleshooting 13 Unable to Change Therapy Mode or Adjust Comfort Control Nursing Procedures 13 14 Recommended Linen Changing the Therapy Pad Patient Pos...

Page 5: ...ssure relief by combining low air loss with pulsation Low air loss therapy has been demonstrated to reduce the risk of pressure ulcers as well as being a valuable aid in the treatment of pressure ulce...

Page 6: ...for these patients should be considered on an individual basis and discussed with the attending physician The Arise LAL is comprised of two components Therapy control unit Therapy mattress system per...

Page 7: ...ed low pressure heel section provides lower interface pressures Durable base tub is constructed from 100 heavy weight 1680 denier nylon with a 1 5 oz urethane coating and incorporates bed attachment l...

Page 8: ...poor installation of a ground wire that causes the loss of proper ground at the outlet Inadequate grounding at electrical outlets can occur even if there is a ground wire Wires can become loose over t...

Page 9: ...eat section securely to the sleep deck of the bed frame If the therapy pad is not already on the overlay attach it securely using the zippers Hang the control unit on the foot of the bed facing away f...

Page 10: ...setting Refer to Comfort Adjust Setup section Figure 2 on the lower right of the control CairRails by moving the control knob to the up If the patient needs the extended width functionality of the be...

Page 11: ...d Pulsate Static The unit starts in the Static mode which is the standard low air loss therapy Pulsate Pulsate will slightly decrease the pressure in all cushions every 30 seconds then return to the p...

Page 12: ...nd visual alarm Power Interruption Brown out In the event of a brief power interruption up to twenty 20 minutes the Arise LAL control unit will automatically restart and reset to the previous Comfort...

Page 13: ...ly Special linens are not necessary for the Arise LAL While there is no need for a bottom sheet the therapy pad should be covering the therapy cells at all times The patient should never be lying dire...

Page 14: ...should always be in the lowest practical position when the patient is unattended Make sure areas under and around the frame are clear of objects persons and parts of body before adjusting height Warni...

Page 15: ...y pad can be laundered in a washer and dryer with warm water no more than 120 Fahrenheit A non bleach detergent should be used sparingly Wipe dry or allow to air dry Do not use any steam cleaning devi...

Page 16: ...LAL 1000 Mattress 48 122 cm W 80 203 cm H 8 5 22 cm D Arise LAL 1000EX Mattress 36 42 48 91 107 122 cm W 80 88 203 224 cm L 8 5 22cm D USA 120V AC 60 Hz 4A less than 1A running Operating Conditions A...

Page 17: ...Electronic Copy Reference Only Notice Document considered current at time of printing Verify revision before use...

Page 18: ...e instructions is not covered Joerns Arise LAL contains various parts that wear from normal use Joerns Healthcare s obligation under it s warranty is limited to supplying replacement parts servicing o...

Reviews: