Joerns Support Surface

Arise

®

 LAL Mattress Replacement,  

Arise LAL Overlay, Arise 1000, Arise 1000EX

Arise

®

 1000 EX Mattress Setup

Remove the existing mattress from the bed.
Place the Arise 1000 EX mattress with the hose 
connection at the foot end of the bed and the 
therapy cells facing up. Secure the eight (8) 
straps on the mattress securely to the loops 
provided on the Bari10A bed frame.  

Note:

 These straps should not be over tightened 

as they may interfere with the Bari10A bed deck 
elements during width reduction.
If the therapy pad is not already on the mattress, 
attach it via the longitudinal zippers to the 
mattress. 

Note:

 This will require threading the 

hose set through the control window opening 
on the therapy cover located on the mattress 
side at the patients left foot. Also attach the 
snaps around the control panel opening from 
the therapy cover to the mattress control panel 
in four (4) places so that the control panel is not 
obscured (Figure 2).
Hang the control unit on the foot of the bed 
facing away from the bed. Attach the hose 
connector marked 

CPR

 to the underside of the 

control unit.
Plug the Fowler boost sensor into the side of the 
control unit. 
Plug in the control unit and the yellow 

Standby 

light will illuminate. Press the 

Power 

key.  

The control unit will start and the green light  
will illuminate.
Allow three to five minutes for full inflation
Select the appropriate 

Comfort Adjust

 setting 

(Refer to “Comfort Adjust Setup” section).
The CairRails air bolsters can be inflated or 
deflated as required. Locate the control knob 
(Figure 2) on the lower right of the control 
panel on the mattress. Next, inflate/deflate the 
CairRails by moving the control knob to the up 
(inflate) or down (deflate) position. 
If the patient needs the extended width 
functionality of the bed (36", 42" or 48") then 
complete the following procedure. 

Note:

 this can 

be accomplished with the patient on the bed:

Adjust the Bari10A bed frame with the knee 
section in the flat position.
Widen the bed frame (see bed manual) 
Un-strap the mattress side straps from the 
Bari10A bed frame. 
Un-snap the pocket that contains the 
extension side cells so they are free  
to inflate.



1.

2.
3.

4.

Attach the extended width mattress side 
straps to the extended bed deck (

Note:

 do not 

over tighten).
Inflate the extension side cells by turning  
the lower left knob on the mattress control 
panel up. After 2-5 minutes the cells will 
inflate. Leave the knob in this position to 
maintain inflation.

If the patient needs the extended length 
functionality of the bed (80" - 88" then complete 
the following procedure. 

Note:

 this can be 

accomplished with the patient on the bed:

Adjust the Bari10A bed frame with the knee 
section in the flat position.
Extend the bed frame (see bed manual) 
Un-strap the mattress foot straps from the 
Bari10A bed frame. 
Un-snap the pocket that contains the 
extension foot cells so they are free  
to inflate.
Attach the extended length mattress foot 
straps to the extended bed deck (

Note:

 do 

not over tighten).
Inflate the extension foot cells by turning the 
lower left knob on the mattress control panel 
(Figure 2) up. After 2-5 minutes the cells will 
inflate. Leave the knob in this position to 
maintain inflation. 

5.

6.

1.

2.
3.

4.

5.

6.

10

Figure 2

Electronic Copy: Reference Only

Summary of Contents for arise1000

Page 1: ...r Service Manual Joerns Support Surface Arise LAL Mattress Replacement Arise LAL Overlay Arise 1000 Arise 1000EX transforming wound management To avoid injury read user s manual before using Electroni...

Page 2: ...stem always ensure that the patient is positioned properly within the confines of the bed Do not let any extremities protrude over the side or between the bed rails when the mattress is being used Dan...

Page 3: ...aps and spaces between bed system components and clinical guidance for assessment and implementation of bed side rails in various health care settings Entrapment zones involve the relationship of bed...

Page 4: ...and Monitoring Additional Features 12 13 CPR Transport Power Failure CairRails Optional Accessories Troubleshooting 13 Therapy Surface Not Inflating Unable to Change Therapy Mode or Adjust Comfort Con...

Page 5: ...from these damaging forces 5 Introduction The Arise provided by Joerns Healthcare is a unique therapy system that provides pressure relief by combining low air loss with pulsation Low air loss therap...

Page 6: ...l times The use of the Arise for these patients should be considered on an individual basis and discussed with the attending physician Features The Arise is comprised of two components Therapy control...

Page 7: ...ut for most patients within weight limit Integrated low pressure heel section provides lower interface pressures Durable base tub is constructed from 100 heavy weight 1680 denier nylon with a 1 5 oz u...

Page 8: ...proper ground at the outlet Inadequate grounding at electrical outlets can occur even if there is a ground wire Wires can become loose over time at the connection to the outlet Note To install new wir...

Page 9: ...lls facing up Secure the straps around the head and foot ends of the mattress Fasten the two straps by the seat section securely to the sleep deck of the bed frame 4 5 If the therapy pad is not alread...

Page 10: ...flated as required Locate the control knob Figure 2 on the lower right of the control panel on the mattress Next inflate deflate the CairRails by moving the control knob to the up inflate or down defl...

Page 11: ...urn to the mode it was in prior to Autofirm either therapy or pulsate in approximately 12 minutes The Arise has two low air loss therapy modes Therapy and Pulsate Therapy The unit starts in the Therap...

Page 12: ...to twenty 20 minutes The power interruption feature keeps the current settings in memory based on the frequency of use and level of battery charge The Alarm Reset will turn off audible and visual alar...

Page 13: ...ke sure the Lockout function is disabled To disable press the up and down comfort adjust arrows simultaneously 1 2 3 4 Nursing Procedures Recommended Linen Special linens are not necessary for the Ari...

Page 14: ...s Early intervention may be essential to preventing serious skin breakdown Bed Height To minimize the risks of falls or injury the patient surface should always be in the lowest practical position whe...

Page 15: ...warm water no more than 120 Fahrenheit A non bleach detergent should be used sparingly Wipe dry or allow to air dry Steam Cleaning Do not use any steam cleaning device on the unit Excessive moisture...

Page 16: ...8 122 cm W 80 203 cm H 8 5 22 cm D Arise 1000EX Mattress 36 42 48 91 107 122 cm W 80 88 203 224 cm L 8 5 22cm D Electrical Specifications USA 120V AC 60 Hz 4A less than 1A running Environmental Condit...

Page 17: ...www joerns com info joerns com Joerns Support Surface Arise LAL Mattress Replacement Arise LAL Overlay Arise 1000 Arise 1000EX 17 Notes Electronic Copy Reference Only...

Page 18: ...www joerns com info joerns com Joerns Support Surface Arise LAL Mattress Replacement Arise LAL Overlay Arise 1000 Arise 1000EX 18 Notes Electronic Copy Reference Only...

Page 19: ...www joerns com info joerns com Joerns Support Surface Arise LAL Mattress Replacement Arise LAL Overlay Arise 1000 Arise 1000EX 19 Notes Electronic Copy Reference Only...

Page 20: ...authorized in writing by Joerns Healthcare will void this warranty Damage caused by use in unsuitable environmental conditions abuse or failure to maintain the product in accordance with user and serv...

Reviews: