JK2830 Series Operation Manual Chapter 3 Introduction to [DISP]
38
3.6.7
Open/Short correction on <MEAS DISP> page
Operation steps of open correction function
1)
On <MEAS DISP> page, press the key
[CLR]
, the following soft keys will be
displayed:
◼
SPOT OPEN
◼
SWEEP OPEN
◼
OPEN OFF ON
◼
MORE 1/2
2)
Connect test fixture to test terminal. The fixture is open and not connecting to
any DUT.
3)
Press
SPOT OPEN
, JK2830 will make open correction on the current
frequency.
4)
Press
SWEEP OPEN
, JK2830 will test the open admittance (capacitance and
inductance) under 37 frequencies. It will take about 20 seconds to finish the
open full-frequency correction. In the process of correction , the following
soft key will be displayed:
◼
CANCEL
This soft key can be used to terminate the current open correction operation
and reserve the formal open correction data.
5)
Press
OPEN OFF ON
to turn ON/OFF the function of open-circuit correction,
If ON, JK2830 will perform open-circuit correction calculation in the later
testing process. If OFF, no open correction calculation will be taken in later
measurement.
Operation steps of short correction function
1)
On <MEAS DISP> page, press the key
[CLR]
, the following soft keys will be
displayed:
◼
SPOT OPEN
◼
SWEEP OPEN