22/25
Chapter 4 Measurement
This chapter describes the functions for correct measurement in stages, including the start test
phase, test phase, test display phase, test completion phase, and discharge phase.
4.1 Start the test
1. Set the relevant parameters
2. Press the [START] key to trigger the test
3. The test starts, the output voltage of the test terminal, the [STOP] button flashes.
Trigger mode meaning
Manual trigger Press [START] manually to trigger the test
External IO trigger Trigger test by external EXT.IO terminal START signal
Trigger by external command Trigger test by RS232, LAN port command
EXT.SW port trigger Test by external trigger switch on the panel
note:
• You cannot restart another test while the test is not over.
• When the STOP signal of the EX.I / O port is LOW, the test cannot be triggered
• When the InterLock signal of the EX.I / O port is LOW, the test cannot be triggered
4.2 Measured value display
The following is the test range. Once it exceeds the following range, OVER.F and UNDER.F are
displayed.
Test voltage and range:
Test voltage
Resistance range
Display range (
Ω
)
Resolution (
Ω
)
25V
≤
V
<
100V
2M
Ω
0.000~4.000M
0.001M
20M
Ω
1.90M~40.00M
0.01M
200M
Ω
19.0M~400.0M
0.1M
100V
≤
V
<
500V
2M
Ω
0.000~4.000M
0.001M
20M
Ω
1.90M~40.00M
0.01M
200M
Ω
19.0M~400.0M
0.1M
2000M
Ω
190M~4000M
1M
500V
≤
V
≤
1000V
2M
Ω
0.000~4.000M
0.001M
20M
Ω
1.90M~40.00M
0.01M
200M
Ω
19.0M~400.0M
0.1M
4000M
Ω
190M~9990M
1M
4.3 Test termination
There are two ways to terminate the test, one is the forced termination method, and the other is the
automatic termination method.
The forced mode can be terminated at any stage of the test when a forced termination test
instruction or signal is encountered.