Adjustment of anti-tipper
Adjust anti-tipper angle by unscrew the nut (picture 5).
Picture 5
Picture 6
Use of Hand Parking Brakes
Push hard the manual brake forward to stop the rear wheel, then the wheelchair will stop moving.
Release the handbrake, wheelchairs will be back to normal driving status. The users can select
manual mode or power mode according to their needs (picture 6).
Switch between Manual Mode and Electric Mode
Manual mode: turn off power and brake the wheelchair first, pull the clutch (red part) out of the groove
and spin 45°. ( Picture 7-1)
Electric mode: spin the clutch (red part) and put it back to groove, lightly push the wheelchair to hear a
click sound. ( Picture 7-2)
Picture 7-1
Picture 7-2
When changing from manual mode to electric mode, please confirm both side clutches are
inside the grooves. Any side clutch is not fully into groove, may cause danger of rollover.
Do not make wheelchair into manual mode when user is driving on slopes.
4 Battery and Charger
4.1 Charger Requirements
The charger is used for battery charging. During electric wheelchair is charging, do not
use it.
Charger Technical Data
Input voltage: AC 220V (110V)
±
10%
Output voltage: 24V/DC
Output current: 2A/4A
4.2 Usage of Charger
In order to charge the battery, connect the charger with power supply and battery box’s plug.
Do as following instructions to complete charging process:
Step 1
: Make sure the charger
groove is not blocked.
Step 2:
Make sure the electric wheelchair is power off.
Step 3:
Unplug the plugs which connect the battery box and control.
Step 4:
Connect the charger’s output plug to the battery box’s power plug.
Step 5:
Connect the charger’s main plug to the power supply and
the red
light will be alight. Fully charging needs 8 to 10 hours, do
not overcharge more than 24 hours.
Ste
p 6: As to prevent cut short battery life, please charge the battery at least once per month
when
the wheelchair is not being used.