Image observation after sample loading: 
 

1.

 

Select desired signals by click “Condition” icon in the icon window 1.  

 
2.

 

Setting  the  SpotSize  (SS)  display  in  the  active  data  display  by  clicking  one  of  the 
present button or select the desired Spotsize by dragging the slide button. For a normal 
image, set the spot size SS about 30. For high-resolution observation, set the spot size 
SS < 30. When perform analysis and other operations, set the spot size SS > 30.  

 

3.

 

Adjust  the  image  by  using  automatic  function  icons  “AF”,  “AS”  and  “ACB”  in  the 
icon  window  1.    If  click  “AF”,  the  image  focus  is  automatically  adjusted.  If  click 
“ACB”, the image contrast and the image brightness are automatically adjusted.  

 

4.

 

Set  the  desired  magnification.  [Focus],  [Stigma],  [Contrast  and  Brightness]  can  be 
adjusted  with  the  adjustment  wheel  in  the  icon  window.  When  you  want  to  switch 
[Focus], [Stigma], [Contrast and Brightness], click on the button of the upper part of 
the wheel. For details, please refer to the quick guide page 8. We can also adjust Focus, 
Stigma, Contrast and Brightness by using the keyboard.  

 
5.

 

Selecting the scan speed according to needs.  “FAST” selects the scan rate in the fast 
mode and is suitable for observing an image. “SLOW” selects the scan rate in the slow 
mode  and  is  suitable  for  observing  the  detail  of  the  image.  “Freeze”  stops  scanning 
and  freeze  an  observing  image.  To  cancel  freeze,  select  any  one  of  the  scan  icons. 
“STORE” selects the scan rate in the store mode, which is used to acquire a checked 
image and save the image automatically.  

 