3 SETTING OF INSTRUMENT CONDITIONS
3.6 REFERENCE SETTING
In the solid-state NMR, the reference cannot be corrected by using the lock signal as in
the solution NMR. Therefore, the reference is set up using the spectrum of an exter-
nal-reference sample. There are two methods for setting the reference: setting in the Delta
and setting using the Z0 shim.
3.6.1 External Reference Setting
1.
Prepare the spectrum with the external-reference substance, and carry out
the reference setting.
Refer to the user’s manual “Data processing” for details for setting the refer-
ence.
2.
In the
Delta Console
window, select
Tools
-
Geometry Tools
-
Copy Ruler
Reference
.
Fig. 3.6 Delta Console window
The
Copy Ruler Reference
window opens.
3.
First, click the
Source
button, and select the spectrum with the external ref-
erence. Next, click the
Target
button, and select the spectrum on which you
want to set the reference.
Fig. 3.7 Copy Ruler Reference window
Hidden buttons, such as
X
, appear.
4.
Click an axis to set the reference.
The reference is set up.
3-8