To change the maximum absorbance or % transmittance
value press the key adjacent to the maximum y-axis
value, this opens a number entry screen. Use the keys
adjacent to the
arrow
icons to increase or decrease the
number. To change the sign press the key below the
+
or
-
icon. Repeat presses will cycle between
+
and
–
. Once
the required absorbance or % transmittance value has
been entered press the key adjacent to the
tick
icon.
7.3
CALIBRATION
In the spectrum measurement mode the calibration is
a baseline scan which is performed across the selected
wavelength range at the selected scan interval. Insert a
cuvette containing the blank solution into the sample
chamber and close the instrument lid. Press the key below
the
baseline scan
icon to initiate the baseline scan. The
baseline icon will change to show
baseline scan in
progress
icon and a progress bar will be displayed.
To stop the baseline scan before completion press the key
below the
baseline scan in progress
icon. Confirmation
will be needed to stop the baseline scan. Press the key
adjacent to the
tick
icon to confirm stopping the baseline
scan and return to the expanded operating menu.
Press the key adjacent to the
cross
icon to continue the
baseline scan. Once the baseline scan has been completed
the
scan sample
icon is displayed and a sample can be
measured.
If the wavelength range, or the scan interval, is changed
before a sample scan is performed a new baseline scan
must be performed across the new wavelength range, at
the new scan interval, before a sample can be measured.
7.4
SAMPLE MEASUREMENT
It is not possible to measure a sample before a baseline
scan has been performed. Insert a cuvette containing the
sample to be measured in the sample chamber and close
the instrument lid. Press the key below the
scan sample
icon to start the spectrum scan of the sample.
The instrument will perform a scan across the wavelength
range and scan interval previously selected. The
scan
sample
icon will change to show the
spectrum scan is
in progress
icon.
09:02
2.500
ABS
-0.500
350
550
450nm
0
100
09:02
2.500
ABS
-0.500
350
550
450nm
09:02
2.500
ABS
-0.500
350
550
450nm
09:02
2.500
ABS
-0.500
350
550nm
450nm
3
+
.
0
2
0
Summary of Contents for 7315
Page 1: ...Spectrophotometers Models 7310 7315 Operating Manual 731 005 REV D 06 10...
Page 2: ......
Page 74: ...SECTION 14 Declaration of Conformity 76...
Page 75: ...SECTION 14 Declaration of Conformity 77...
Page 85: ......