SECTION 5 - Care and Maintenance

Over time the electrode sensor will degrade, but regular calibrating, cleaning and 
the correct storage will prolong its life. The epoxy bodied combination electrode 
supplied with this instrument is suitable for the majority of tests carried out in 
aqueous solutions. For other applications, such as low ionic strength, Tris buffers, 
high  temperatures  and  strongly  acidic  solutions  a  more  suitable  pH/reference 
electrode  pair  may  be  required.  Details  or  advice  are  supplied  on  request 
([email protected]). 

If  the  electrode  readings  are  slow  or  erratic,  place  the  electrode  into  cleaning 
solution or 7.00 pH solution for at least half an hour to one hour before testing 
again. The temperature probe and pH electrodes are not waterproof and cannot 
be fully immersed in water.  It is recommended that the instrument is calibrated 

daily to achieve consistent and accurate results. 

The following general guidelines indicate the care and maintenance required:

Always rinse the pH electrode with cleaning solution or deionised water before 
next use.

Electrodes should be stored:

● 

away from direct sunlight

● 

in a vertical position

● 

within their specified temperature range

Always ensure the electrode is used within its specified temperature range. Ageing 
of electrodes used above their specified temperature is rapid and irreversible.

DO NOT

 touch the sensitive glass pH membrane or reference junction. Excess 

droplets of solution may be removed by gently blotting with filter paper or tissue.

DO NOT

 rub the electrode as this may induce an electrostatic charge.

During use ensure the electrode is rinsed between each measurement to eliminate 
the contamination of solutions.

15

Summary of Contents for 570

Page 1: ...pH mV and Temperature Meter Model 570 Instruction Manual 570 350 Rev B 05 17...

Page 2: ......

Page 3: ...f chemicals should be employed 5 If it is suspected that safety protection has been impaired in any way the unit must be made inoperative and secured against any intended operation The fault condition...

Page 4: ...nweisungen in diesem Handbuch ein sicheres Arbeitssystem einsetzen 3 Mit Ausnahme der Teile deren Wartungsverfahren in diesem Handbuch beschrieben sind enth lt dieses Ger t keine weiteren Teile die vo...

Page 5: ...ar necessario pagare un importo per le riparazioni effettuate 4 sempre necessario fare riferimento ai dati sulla salute e sulla sicurezza forniti con le sostanze chimiche utilizzate Tutte le informazi...

Page 6: ...suministran con cualquier producto qu mico que se utilice Deben respetarse atentamente todas las indicaciones recomendaciones y advertencias disponibles en relaci n con la manipulaci n el almacenamien...

Page 7: ...mperature Adjustment 12 3 1 MANUAL TEMPERATURE ADJUSTMENT 12 3 2 AUTOMATIC TEMPERATURE CORRECTION ATC 12 SECTION 4 Calibration and Measurement 13 SECTION 5 Care and Maintenance 15 5 1 BATTERY REPLACEM...

Page 8: ...Range 2 to 16pH pH Resolution 0 01pH pH Accuracy 0 02pH mV Range 1000 to 1000mV mV Resolution 1mV mV Accuracy 1mV ATC 0 to 100 C Temperature Range 39 9 to 149 9 C Temperature Resolution 0 1 C Tempera...

Page 9: ...ture probe 170 101 pH 4 7 and 10 buffer sachets 3 x AAA alkaline batteries fitted Instruction manual 570 350 Jenway Foreign Manual CD JENMANCD 2 2 INSTALLATION Connect the electrode to the BNC socket...

Page 10: ...showing pH or electrode potential and temperature reading 2 Connector for temperature probe 3 BNC connector for pH electrode 4 UP arrow Key for adjusting values 5 DOWN arrow Key for adjusting values...

Page 11: ...instrument temperature units between C and F press and hold the MODE key for 3 seconds When the instrument is turned on a single press of the CAL and backlight key will turn the backlight on Another p...

Page 12: ...degrees Note that manual temperature adjustment can only be performed if the temperature probe is not connected 3 2 AUTOMATIC TEMPERATURE CORRECTION ATC When the temperature probe is connected the 570...

Page 13: ...measured using a thermometer so that the temperature can be manually adjusted Place the pH electrode and temperature probe into 7 00 pH buffer solution and allow the reading to stabilise Stir then pre...

Page 14: ...perature of 25 C and leave for 30 seconds to stabilise Record the mV reading Asymmetry Potential Rinse the electrode and temperature probe in distilled or deionised water blot and repeat the process i...

Page 15: ...s are not waterproof and cannot be fully immersed in water It is recommended that the instrument is calibrated daily to achieve consistent and accurate results The following general guidelines indicat...

Page 16: ...ery icon is low This meter will continue to measure accurately but after further usage the meter will display FLAt bAT and shutdown Unscrew the screws on the back of the meter and replace with three A...

Page 17: ...sachets pack of 10 025 180 pH 7 buffer sachets pack of 10 025 181 pH 10 buffer sachets pack of 10 025 192 pH electrode storage solution 250ml 025 161 pH electrode cleaning solution 500ml 570 401 White...

Page 18: ...clear description of the fault and a completed copy of our Decontamination Certificate Please clearly mark the package for the attention of the Service Department and post to the following address Col...

Page 19: ...anufacturer Product Laboratory Equipment Manufacturer Cole Parmer Ltd Beacon Road Stone Staffordshire ST15 0SA United Kingdom File Number P225 Authorised Cole Parmer Representative Beacon Road Stone S...

Page 20: ...Cole Parmer Ltd Beacon Road Stone Staffordshire ST15 0SA United Kingdom Tel 44 0 1785 812121 Email cpinfo coleparmer com Web www jenway com...

Reviews: