3.6.5 G.L.P Setup …
Exit
enables the user to exit the Setup menu and return to the main
set up screen.
Cal Reminder
can be enabled or disabled by using the Up/Down keys which toggle
between the two settings.
Cal Reminder Interval
can be set within the limits of 001 to 999 hours by using the Up/Down
keys.The Left/Right arrow keys allow forward and backward
movement along the row of digits. Press the STO key to accept the
value.
Cal Reminder Audible Alarm can be enabled or disabled by using the Up/Down keys which toggle
between the two settings.
User ID
up to a 4 digit code can be set using the Up/Down keys. The Left/
Right arrow keys allow forward and backward movement along the
row of digits. Press the STO key to accept the value
3.6.4 Alarms Setup...
Exit
enables the user to exit the Setup menu and return to the main
set up screen.
Alarm Outputs
can be enabled or disabled by using the Up/Down keys which toggle
between the two settings.
Audible Alarm Warning
can be enabled or disabled by using the Up/Down keys which toggle
between the two settings.
pH Alarm High
allows the user to set the high alarm limit up to 20.000 pH.
pH Alarm Low
allows the user to set the low alarm limit down to -2.000 pH.
mV Alarm High
allows the user to set the high alarm limit up to +1999.9mV.
mV Alarm Low
allows the user to set the low alarm limit -1999.9 mV.