background image

13

351 050/REV B/09-06

7

The voltage developed by each individual pH electrode in the presence of a known hydrogen ion

concentration is theoretically predictable, but in practise deviations from the theoretical value can be

expected. These deviations will change slowly during the life of an electrode. It is therefore essential

to routinely calibrate the system using solutions with a known and constant pH value. These

solutions are called buffers.

3.3 Preparation of Buffer Solutions

Care must be taken in the preparation of all buffer solutions. The correct quantity of distilled or

deionised water should be used when preparing the solutions. For accurate and repeatable results

it is essential to follow the manufacturers instructions carefully.

3.4 Solution Temperature Values

The value of all buffer solutions varies with solution temperature. For accurate calibration of

electrodes using buffer solutions, it is necessary to measure the temperature of the buffer solution

being used. The unit should then be calibrated to the corrected pH value. Manufacturers of buffer

powders and solutions will provide a table of values at varying temperatures for their buffers.

Note:

Buffer solutions will contaminate with exposure to air and should be stored

in airtight containers when not in use. Used solution should be discarded and

not returned to the container as this will cause contamination.

For best results fresh solutions should be prepared prior to calibration.

Summary of Contents for 3510

Page 1: ...Instruction Manual 351 050 REV C 02 17 pH mV Temperature Meter Model 3510...

Page 2: ......

Page 3: ...to the appropriate servicing authority Merci de lire attentivement ces informations avant d installer ou d utiliser cet appareil 1 L appareil d crit dans ce manuel est con u pour tre utilis uniquemen...

Page 4: ...irgendeiner Weise besch digt wurden muss das Ger t au er Betrieb genommen und gegen weiteren Gebrauch gesichert werden Die St rung sollte der zust ndigen Serviceeinrichtung unverz glich gemeldet werd...

Page 5: ...adas que se especifican en este manual 3 Cualquier elemento que no se encuentre entre los definidos en los procedimientos de mantenimiento aqu descritos no podr utilizarse en este instrumento La extra...

Page 6: ...Solution Temperature Values 3 4 Good Practice Guidelines 3 5 Instrument Set Up 3 6 pH Calibration 3 7 Error Codes 3 8 mV Mode 3 9 Performing Measurements 3 10 Results Storage and Display 3 11 Section...

Page 7: ...Range 2 000 to 19 999pH Resolution 0 001 0 01 0 1pH Accuracy 0 003pH mV Absolute or Relative Range 1999 to 1999mV Resolution 0 1mV Accuracy 0 2mV Input Impedance 1012 ohms Temperature Measuring Ranges...

Page 8: ...3 ATC probe 027 500 4 Electrode holder 903 300 5 4 7 and 10pH buffer sachets 6 BNC shorting plug 009 146 7 Power Supply as specified at time of ordering the product 8 Condensed operating instructions...

Page 9: ...sed 6 Endpoint symbol this symbol is displayed when the pH changes by less than 0 005pH 0 2mV over a five second period Once an endpoint has been detected the reading must change by more than 0 005pH...

Page 10: ...in mV mode 3 Print key used to initiate a print 4 Up Arrow used for adjustment during set up to scroll results and to toggle between mV and pH modes 5 Down Arrow used for adjustment during set up to s...

Page 11: ...e is needed 2 pH Socket BNC type socket which allows combination pH or redox electrodes to be used 3 Temp Socket 8 pin mini DIN socket This allows the Automatic Temperature Compensation ATC probe 027...

Page 12: ...dicators change colour as the pH level varies These can be used as a guide to the pH level but can be limited in accuracy and difficult to interpret correctly in murky or coloured samples For greater...

Page 13: ...he solutions For accurate and repeatable results it is essential to follow the manufacturers instructions carefully 3 4 Solution Temperature Values The value of all buffer solutions varies with soluti...

Page 14: ...n use Note Some epoxy gel filled electrodes are not refillable 5 DO NOT touch the sensitive glass pH membrane or reference junction during use Excess droplets of solution may be removed by gently blot...

Page 15: ...parameters not saved will remain as defaults or previous setting 3 6 1 Number of calibration points entry I 2 or 3 point calibration is possible on the 3510 To set these parameters Select SETUP mode...

Page 16: ...keys are used to select the next adjusting digit which will flash when selected Once the full reading has been adjusted to the correct value press the STO key If a 1 point calibration only is being pe...

Page 17: ...olution press the STO key The symbol will disappear and the secondary display will momentarily show SAVED and the next set up menu option will be displayed To exit the set up menus at any time press t...

Page 18: ...e MODE menu 3 6 5 Manual temperature If the ATC probe is not connected to the instrument the display will show The secondary display will show MAN C or F and will then scroll MANUAL TEMPERATURE C or F...

Page 19: ...no function during this set up Symbol will flash while adjustment is being made To accept the selected baud rate press the STO key The symbol will disappear and the secondary display will momentarily...

Page 20: ...ure compensation is being used the annunciator will indicate MAN The status display will indicate the slope value 2 CAL 1 Immerse the electrode s in the first buffer solution and allow the instrument...

Page 21: ...he endpoint symbol will be displayed Press the CAL or STO key 15 351 050 REV B 09 06 If a successful I point calibration has been performed the secondary display will momentarily show CAL OK The instr...

Page 22: ...V B 09 06 16 If a successful I point calibration has been performed the secondary display will momentarily show CAL OK The instrument will then update the display The display will then show the next p...

Page 23: ...he instrument will then update the display Rinse the electrode s in deionised water If the instrument fails a calibration the error message EO will be shown and ERROR SLOPE OUTSIDE LIMITS will scroll...

Page 24: ...at corrective action should be taken Recording these values as part of a routine Quality Control program can give a good indication of the condition of the electrode 351 050 REV B 09 06 Relative Milli...

Page 25: ...ompensation is being used the preferred measurement range should be selected via the Set Up menu b Immerse the electrode into the solution and set the to the solution temperature via the Set Up menu 4...

Page 26: ...V is selected Selecting the RESULTS menu displays the reading stored at the displayed memory location The Up arrow increments the memory index the Down arrow decrements the memory index The Left arrow...

Page 27: ...tand in the solution until clean Rinse and repeat as necessary Electrodes which have been allowed to dry out should be soaked overnight in warm distilled water Table of Cleaning Agents for Glass Elect...

Page 28: ...Power supply for printer M7817X1 US Power supply for printer 037 801 Interface cable kit 33 Section 5 Optional Accessories 5 1 Optional Accessories The following list of items are available as optiona...

Page 29: ...interface is available on the rear panel 9 way D type connector The connections are as follows DCD 1 LINKED TO DTR AND DSR RXD 2 INPUT TO 3510 TXD 3 OUTPUT FROM 3510 DTR 4 LINKED TO DCD AND DSR GND 5...

Page 30: ...ntrol and software XON XOFF flow control Pressing the PRINT key outputs from the RS232 interface Sending an ASCII D to the 3510 causes a printout of the current displayed reading plus sample number Se...

Page 31: ...Most recent calibration information Eo value Slope efficiency Buffer type This will be followed by results data in either pH or mV dependent on mode selected Details will also be given on temperature...

Page 32: ...32 6 4 1 Example Printout 3510 Header printout 3510 Results printout 26 351 050 REV B 09 06...

Page 33: ...0 see 7 2 Replace electrode Unable to recognise Using correct buffer set Is the buffer type correct Use AUTO for buffers Jenway supplied buffers Replace buffers Use BNC cap to test 3510 see 7 2 Replac...

Page 34: ...r panel socket The secondary display will momentarily show E2 RST 4 If this does not resolve the problem please contact the manufacturer or your local distributor 7 2 Functional check The measurement...

Page 35: ...of the above manufacturer Product Laboratory Equipment Manufacturer Cole Parmer Ltd Beacon Road Stone Staffordshire ST15 0SA United Kingdom File Number P225 Authorised Cole Parmer Representative Beaco...

Page 36: ...Cole Parmer Ltd Beacon Road Stone Staffordshire ST15 0SA United Kingdom Tel 44 0 1785 812121 Email cpinfo coleparmer com Web www jenway com...

Reviews: