JENCO VisionPlus pH618N Operation Manual Download Page 6

 

Key

 

Operation
Mode

 

Duration

 

Function

 

CAL

 

Recall Data

2 seconds

Erases ALL stored 
data and return to 
Measure Mode.

 

 

Measure

 

 

2 seconds

Stores current pH 
reading, date and 
time into memory.

Date and 
Time Set

 

 

0 second

 

Increases current 
number.

 

Recall Data

2 seconds 
or more

Leaves Recall Data
Mode.

 

 

 
 
 

STORE

 

Buffer Select 2 seconds 

or more

Leaves Buffer 
Select Mode.

 

 

Measure

 

 

2 seconds

Enters Recall Data
Mode.

 

 

Measure

 

5 seconds 
or more

Enters Buffer 
Select Mode.

 

 

Recall Data

 

0 second

 

Scrolls through 
stored data.

Date and 
Time Set

 

 

0 second

 

Decreases current 
number.

 
 
 

 

SCROLL

 

Buffer 
Select

 

 

0 second

 

Selects new buffer 
set.

 

USING VISIONPLUS pH 618N   

 

 

A.  Power ON/OFF

 

Press “HOLD” key for 2 seconds or more. The unit will turn on 
and enter the “Measure Mode”. Repeat the process to turn off 
the unit. The unit will also automatically turn off after 10 minutes 
of no key activity. 

 

B.    Set Date and Time

 

1. Press “CAL” key for 10 

seconds or more to enter 
“Date and Time Set Mode”.

 

 

 

2. Press “HOLD” key to select 

calendar or time parameters. 

 

3. Press “STORE” and 
“SCROLL” keys to increase and decrease values respectively. 

 

4. Press “CAL” key for 2 seconds to exit and return to 

“Measure Mode”. 

 

 

Summary of Contents for VisionPlus pH618N

Page 1: ...Operation Manual MODEL pH 618N Microcomputer Based pH Temperature Pocket Meter pH618N...

Page 2: ...ption 3 B VisionPlus Graphical Display 3 OPERATION MODES AND KEYPAD OPERATIONS 4 A Operation Modes 4 B Keypad Operations 4 USING VISIONPLUS pH 618N 5 A Power ON OFF 5 B Set Date and Time 5 C Select Bu...

Page 3: ...nstrument can operate within specifications for approximately2 3 hours after low battery indicator appears 1 Shown in the right figure 2 Take off the battery cover 3 Remove all of the old batteries an...

Page 4: ...r 5 Electrode ATC assembly Electrode cap is not shown B VisionPlus Graphical Display 1 DATE MM DD YY 2 HOLD mode indicator 3 MEASURE mode indicator 4 CALIBRATION buffer s 5 TIME HH MM 24H format 6 pH...

Page 5: ...et Mode is used to change date and time 6 Buffer Select Mode Buffer Select Mode is used to select the buffer set which can either be 7 00 7 00 4 01 10 01 or 6 86 6 86 4 00 9 18 B Keypad Operations Key...

Page 6: ...d Scrolls through stored data Date and Time Set 0 second Decreases current number SCROLL Buffer Select 0 second Selects new buffer set USING VISIONPLUS pH 618N A Power ON OFF Press HOLD key for 2 seco...

Page 7: ...H meter will perform end point sensing to determine when the calibration reading is stable Meter is calibrated when WAIT icon disappears and the unit will automatically enter the next point calibratio...

Page 8: ...E and M XX icons will appear indicating the reading has been saved and stored in memory location XX The unit will automatically returns to Measure Mode and ready to perform other task Note The Model p...

Page 9: ...trode ATC assembly as shown in the right figure 2 Remove the old electrode from the electrode collar 3 Insert a new electrode make sure the electrode fit back into the meter correctly 4 Screw back the...

Page 10: ...0 00 to14 00pH range Over Range or Recalibrate OVER 0 0 60 0 C pH CAL a pH STAND b pH SLOPE a Offset 7 00pH mV 90mV Offset 6 86pH mV 98 3mV b New slope ideal slope by 30 Use a new buffer solution Repl...

Page 11: ...Electrode Slope recognition 30 at pH 4 00 4 01 9 18 and 10 01 Input impedance 1012 Temperature Temperature sensor Thermistor 10 k at 25 C Temperature unit Default in factory General Power Meter LR44...

Page 12: ...protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss Jenco will not be responsible for damage resulting from careless or insufficient pa...

Page 13: ...618N TC CHEN 2008 7 20 K618N 01 2009 4 3 02 2009 4 3 01 618 MANUAL 618 09 01 14 P2 09 4 3 P2 5 P8 P10 P11 BAT 09 4 15 3M KCL FOR 2 HOURS...

Reviews: