10
SPECIFICATIONS
Display
Range
Resolution
Accuracy
Conductivity/
TDS
K=0.01
0.000 to 1.999 uS/cm /
0.000 to 1.999 mg/L
2.00 to 19.99 uS/cm /
2.00 to 19.99 mg/L
0.001 uS/cm /
0.001 mg/L
0.01 uS/cm /
0.01 mg/L
±0.5% F.S.
Conductivity
TDS
K=0.1
0.00 to 19.99 uS/cm /
0.00 to 19.99 mg/L
2.0 to 199.9 uS/cm /
2.0 to 199.9 mg/L
0.01 uS/cm /
0.01 mg/L
0.1 uS/cm
0.1 mg/L
±0.5% F.S.
Conductivity/
TDS
K=1.0
0.0 to 199.9 uS/cm /
0.0 to 199.9 mg/L
200 to 1999 uS/cm /
200 to 1999 mg/L
2.00 to 19.99 mS/cm /
2.00 to 19.99 g/L
0.1 uS/cm /
0.1 mg/L
1 uS/cm /
1 mg/L
0.01 mS/cm /
0.01 g/L
±0.5% F.S.
Conductivity/
TDS
K=10.0
0 to 1999 uS/cm /
0 to 1999 mg/L
2.00 to 19.99 mS/cm /
2.00 to 19.99 g/L
20.0 to 199.9 mS/cm /
20.0 to 199.9 g/L
1 uS/cm /
1 mg/L
0.01 mS/cm /
0.01 g/L
0.1 mS/cm /
0.1 g/L
±0.5% F.S.
Temperature
0.0 to 100.0 °C
0.1 °C
±0.2 °C
Reference Temperature
15.0 to 25.0 °C
Temperature Coefficient
0.0 to 4.0% default at 1.91%
Cell Constant
0.01; 0.10; 1.00; 10.0
TDS Constant
0.30 to 1.00, default at 0.65
Power
100 ~ 240 VAC adapter
or AAA x 6
Calibration Back-up
EEPROM
Audio Feedback
All Touch Keys
Display
(Conductivity/TDS: Temp)
22 mm : 14.5 mm high LCD
Ambient Temperature Range
0 to 50 °C
Relative Humidity
up to 90%
Case
IP54
Dimensions (W x D x H)
150 x 203 x 72 mm
Weight
504 grams(Batteries included)